Tungsten filled 3D printed field shaping devices for electron beam radiation therapy
Lawrie Skinner1, Benjamin P Fahimian1, Amy S Yu1
1Department of Radiation Oncology, Stanford University, Palo Alto, CA, United States of America.
Purpose:
Electron radiotherapy is a labor-intensive treatment option that is complicated by the need for field shaping blocks. These blocks are typically made from casting Cerrobend alloys containing lead and cadmium. This is a highly toxic process with limited precision. This work aims to provide streamlined and more precise electron radiotherapy by 3D using printing techniques.
Methods:
The 3D printed electron cutout consists of plastic shells filled with 2 mm diameter tungsten ball bearings. Five clinical Cerrobend defined field were compared to the planned fields by measuring the light field edge when mounted in the electron applicator on a linear accelerator. The dose transmitted through the 3D printed and Cerrobend cutouts was measured using an IC profiler ion chamber array with 6 MeV and 16 MeV beams. Dose profiles from the treatment planning system were also compared to the measured dose profiles. Centering and full width half maximum (FWHM) metrics were taken directly from the profiler software.
Results:
The transmission of a 16MeV beam through a 12 mm thick layer of tungsten ball bearings agreed within 1% of a 15 mm thick Cerrobend block (measured with an ion chamber array). The radiation fields shaped by ball bearing filled 3D printed cutout were centered within 0.4 mm of the planned outline, whereas the Cerrobend cutout fields had shift errors of 1-3 mm, and shape errors of 0.5-2 mm. The average shift of Cerrobend cutouts was 2.3 mm compared to the planned fields (n = 5). Beam penumbra of the 3D printed cutouts was found to be equivalent to the 15 mm thick Cerrobend cutout. The beam profiles agreed within 1.2% across the whole 30 cm profile widths.
Conclusions:
This study demonstrates that with a proper quality assurance procedure, 3D-printed cutouts can provide more accurate electron radiotherapy with reduced toxicity compared to traditional Cerrobend methods.
Related Concept Videos
Molecular Shapes
Two regions of electron density in a diatomic...
Biological Effects of Radiation
Molecular Shape and Polarity
VSEPR Theory and the Basic Shapes
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Design Example: Identifying the Locations of Monuments in the Field Using Global Positioning System Device