Comparison of the cyclic fatigue resistance of ProGlider, PathGlider and One G path-finding instruments
Damla Özsu Kırıcı1, Ertuğrul Karataş2, Ahmet Demirhan Uygun3
1Department of Endodontics, Faculty of Dentistry, Akdeniz University, Antalya, Turkey.
Abstract:
. The aim of the present study was to compare the cyclic fatigue resistance of novel nickel titanium rotary pathfinding instruments. . Twenty instruments were selected for each file system. A simulated stainless steel root canal, with a 90° angle of curvature and a curvature radius of 3 mm, was used for cyclic fatigue test of the ProGlider (#16, progressive taper: 0.02‒ 0.085), PathGlider (#15, taper: .03), and One G (#14, taper: .03) instruments. Statistical analyses were performed with oneway ANOVA (P=0.05). Post hoc Tukey tests were used to determine any statistically significant differences between the groups. . The ProGlider instruments exhibited significantly more cyclic fatigue resistance than both PathGlider and One G instruments (P<0.001). One G instruments had significantly more resistance to fracture than PathGlider instruments (P<0.05). . ProGlider instruments had better cyclic fatigue resistance than PathGlider and One G instruments.
Related Concept Videos
Mean free path and Mean free time
Path Between Thermodynamics States
Fatigue
Interference: Path Lengths
Two special sources may be considered when they are in phase. This can be easily achieved by feeding the two sources from the same source. An example would be synchronizing the two speakers by feeding them with the same source, such as the sound waves produced by a tuning fork. This setup ensures that the two sources have the same frequency and are...
Behavior of Gas Molecules: Molecular Diffusion, Mean Free Path, and Effusion
Instrument Transformers