Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000
Jing Wang1, Shengmei Zhao2, Zhenguo Su1
1Department of Laboratory, Binzhou Medical University, Yantai, China.
Background:
To compare the Mindray BC-5180 and Sysmex XN-1000 instruments by analyzing the results of complete blood count in the external quality assessment in Shandong Province in 2018.
Methods:
In the external quality assessment, 10 batches of quality control materials were issued throughout the year. The test items were WBC, RBC, Hb, PLT, and HCT. The laboratories using Mindray BC-5180 and Sysmex XN-1000 were screened, and the results were analyzed by t test, Passing-Bablok regression analysis, and Bland-Altman analysis.
Results:
Thirty-six laboratories using Mindray BC-5180 instruments and thirty-six laboratories using Sysmex XN-1000 instruments were screened, and the average difference between the two instruments results is not significant (P > 0.05, t test). Passing-Bablok regression analysis showed that the 95% confidence interval of the regression equation interception of each test item included 0, and the 95% confidence interval of the slope contained 1, r > 0.98, which showed that the correlation is good. The Bland-Altman analysis showed that both instruments had more than 95% of the points within the 95% consistency limit (WBC97.2%, RBC95.6%, PLT97.2%, Hb96.7%, HCT97.5%). Within the consistency limit, the absolute value of the difference between the Mindray BC-5180 instrument and the Sysmex XN-1000 instrument is WBC 0.14%, RBC 0.26%, PLT 2.7%, and Hb 1.9%. HCT is 0.69%, and the difference is clinically acceptable.
Conclusion:
It can be considered that the two instruments have good correlation and consistency, and the two instruments can replace each other.
More Related Videos
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Development of Analytical Methods
Analyte Adsorption and Distribution
Jung's Analytical Theory
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Graphical and Analytic Representation of Sinusoids
The first step is measuring the peak-to-peak value, which is twice the amplitude of the sinusoid. This provides information about the maximum voltage swing of the waveform.
Secondly, the period and angular frequency are determined. The period is the time taken for one complete cycle of the waveform, while...