In vitro validation of indirect calorimetry device developed for the ICALIC project against mass spectrometry
Taku Oshima1, Yves Marc Dupertuis1, Marta Delsoglio1
1Clinical Nutrition, Geneva University Hospital, Geneva, Switzerland.
Rationale:
Accurate evaluation of the energy needs is required to optimize nutrition support of critically ill patients. Recent evaluations of indirect calorimeters revealed significant differences among the devices available on the market. A new indirect calorimeter (Q-NRG®, Cosmed, Roma, Italy) has been developed by a group of investigators supporting the international calorimetry study initiative (ICALIC) to achieve ultimate accuracy for measuring energy expenditure while being easy to use, and affordable. This study aims to validate the precision and the accuracy of the Q-NRG® in the in-vitro setting, within the clinically relevant range for adults on mechanical ventilation in the ICU. Mass spectrometry is the reference method for the gas composition analysis to evaluate the analytic performances of the Q-NRG®.
Methods:
The accuracy and precision of the O2 and CO2 measurements by the Q-NRG were evaluated by comparing the measurements of known O2 and CO2 gas mixtures with the measurements by the mass spectrometer (Extrel, USA). The accuracy and precision of the Q-NRG® for measurements of VO2 (oxygen consumption) and VCO2 (CO2 production) at clinically relevant ranges (150, 250 and 400 ml/min STPD) were evaluated by measuring simulated gas exchange under mechanically ventilated setting at different FiO2 settings (21-80%), in comparison to the reference measurements by the mass spectrometer-based mixing chamber system.
Results:
The measurements of gas mixtures of predefined O2 and CO2 concentrations by the Q-NRG® were within 2% accuracy versus the mass spectrometer measurements in Passing Bablok regression analysis. In a mechanically ventilated setting of FiO2 from 21 up to 70%, the Q-NRG® measurements of simulated VO2 and VCO2 were within 5% difference of the reference mass spectrometer measurements.
Conclusion:
In vitro evaluation confirms that the accuracy of the Q-NRG® indirect calorimeter is within 5% at oxygen enrichment to 70%; i.e. maximum expected for clinical use. Further recommendations for the clinical use of the Q-NRG® by will be released once the ongoing multi-center study is completed.
More Related Videos
09:18Time-resolved ElectroSpray Ionization Hydrogen-deuterium Exchange Mass Spectrometry for Studying Protein Structure and Dynamics
Published on: April 17, 2017
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
Related Concept Videos
Constant Pressure Calorimetry
In Vitro Drug Release Testing: Overview, Development and Validation
Tandem Mass Spectrometry
Mass Spectrometry: Overview
Mass Spectrometry of Amines
Mass Spectrometry: Isotope Effect