Comparison of different devices to measure the intraocular pressure in thyroid-associated orbitopathy
Aylin Garip Kuebler1, Caroline Wiecha2, Lukas Reznicek3
1Department of Ophthalmology, Ludwig-Maximilians-University, Munich, Germany. Aylin.garip-kuebler@med.uni-muenchen.de.
Purpose:
To evaluate the correlation of the intraocular pressure measurements (IOP) with non-contact tonometer Corvis Scheimpflug technology (Corvis ST), Goldmann applanation tonometry (GAT), ocular response analyzer (ORA), and iCARE rebound tonometer in patients with thyroid-associated orbitopathy (TAO) and eye-healthy subjects (control group).
Methods:
Twenty-nine consecutive patients with TAO (79% female) and 30 eye-healthy subjects (60% female) were included in this prospective, age- and sex-matched study. The IOP measurement with Corvis, ORA, GAT, iCARE, and central corneal thickness (CCT) with Corvis was obtained from all study participants.
Results:
The mean age of the patients was 51 ± 10 years in patients with TAO and 56 ± 13 years in the control group. The mean IOP measurements with GAT, Corvis, ORA, and iCARE were 15.93 ± 4.42 mmHg, 18.10 ± 7.54 mmHg, 18.40 ± 7.93 mmHg, and 16.61 ± 7.96 mmHg in patients with TAO and 14.52 ± 3.02 mmHg, 14.48 ± 3.38 mmHg, 15.29 ± 4.64 mmHg, and 14.13 ± 3.85 mmHg in the control group (P = 0.157, P = 0.004, P = 0.017, and P = 0.176 respectively). The mean CCT was 547.5 ± 39.2 μm in patients with TAO and 560.8 ± 49.8 μm in the control group ( P= 0.261).
Conclusions:
The data collected shows an agreement between the iCARE and GAT IOP measurements in TAO patients and in eye-healthy patients. However, the mean value of IOP measurements with Corvis and ORA was significantly higher in patients with TAO in comparison with the control group (P = 0.044 and P = 0.029 respectively).
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
The Thyroid Gland
The follicles have a central cavity lined by simple cuboidal to squamous epithelial cells called follicular cells. These cells produce the glycoprotein...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.