Performance Evaluation of a Newly Developed MR-Compatible Mobile PET Scanner with Two Detector Layouts
Masao Watanabe1, Yuji Nakamoto2, Ryusuke Nakamoto3
1Department of Diagnostic Imaging and Nuclear Medicine, Graduate School of Medicine, Kyoto University, 54 Shogoinkawahara-cho, Sakyo-Ku, Kyoto, 606-8507, Japan.
Purpose:
A mobile positron emission tomography (PET) scanner called flexible PET (fxPET), designed to fit existing magnetic resonance imaging (MRI) or computed tomography (CT) system, has been developed. The purpose of this study was to evaluate the image quality, lesion detection rate, and quantitative values of fxPET compared with conventional bismuth germanium oxide (BGO)-based PET/CT without time-of-flight capability.
Procedures:
Fifty-nine patients underwent whole-body (WB) PET/CT scans approximately 1 h after injection of 2-deoxy-2-[18F]fluoro-D-glucose, followed by the fxPET scans with detectors located above and below the patients (layout A) and with detectors closer to the patients (layout B). Two readers assessed the image quality using a 4-point grade for each layout and reached a consensus. We evaluated the differences and/or correlations between fxPET and WB PET/CT, including the lesion detection rates, the standardized uptake value (SUV), the metabolic tumor volume (MTV), the total lesion glycolysis (TLG), the tumor-to-normal liver ratio (TLR), and the background liver signal-to-noise ratio (SNR).
Results:
The image quality of layout B was better than layout A (p < 0.0001). Of 184 lesions, the detection rate of layout B was significantly higher than WB PET/CT (p = 0.041), while the detection rate of layout A was comparable to WB PET/CT. The SUVmax/mean/peak were larger, and the MTVs were smaller in fxPET than WB PET/CT, especially in the lesions smaller than 2 cm (p < 0.01). The SUVmax/mean/peak, the MTVs and the TLGs of fxPET had significant positive correlations with WB PET/CT (p < 0.0001). The TLRs were significantly larger (p < 0.0001), but the background SNRs were significantly lower in fxPET than WB PET/CT (p < 0.05).
Conclusions:
The fxPET system yielded reasonable image quality and quantitative accuracy. Bringing the detectors closer to the patient yielded improved results.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...