An evaluation of fluid distribution at the implant site during implant placement by using a computational fluid
Shanshan Hu1, Lunliang Duan2, Qianbing Wan3
1Doctoral student, State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu, PR China.
Statement Of Problem:
Heat reduction during implant site preparation is critical. However, studies that assess fluid distribution at the implant site by using saline irrigation as the cooling method during osteotomies are lacking.
Purpose:
The purpose of this study was to evaluate the effect of various parameters on fluid distribution at the implant site by using a computational fluid dynamics numerical model and thus predict the cooling effect at the drill site.
Material And Methods:
The computational fluid dynamics code Flow-3D was adopted to simulate implant site preparation. A 10-mm-deep implant site was prepared by using a 2.2-mm pilot drill, with 4 °C saline sprayed onto the drill from an external injection hole. Different drilling procedures were performed with irrigation volumes of 20, 40, 60, and 80 mL/min at various drill speeds (600, 800, 1000, 1200 rpm) and feed rates (0.5, 1.0, 1.5, 2.0 mm/s), and the fluid distribution under various circumstances was respectively investigated and compared. Data were analyzed by using 1-way ANOVA or the Friedman test according to the normality of the data distribution (P>.05).
Results:
Below the irrigation volume of 60 mL/min, the saline inside the implant site increased with the irrigation volume (P<.001), but further increase in irrigation volume to 80 mL/min had no significant influence on the fluid distribution (P>.05). The obtained fluid had an inverse relationship with the drill speed under the irrigation volumes 20 and 40 mL/min (P<.001), and deeper areas received less cooling under 20 mL/min (P<.001). However, no significant differences were observed under 60 and 80 mL/min (P>.05). In addition, the variation of feed rate had no significant effect on the mean fluid fraction for all the tested groups (P>.05).
Conclusions:
The fluid distribution at the implant site could be affected by the irrigation volume and drill speed but was not correlated with the feed rate.
More Related Videos
06:21Osmotic Minipump Implantation for Increasing Glucose Concentration in Mouse Cerebrospinal Fluid
Published on: April 7, 2023
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
The Fluid Mosaic Model
Fluid Mosaic Model
Accelerating Fluids
The motion of the liquid within this infinitesimal cylinder is considered to obtain the pressure difference. Three vertical forces act on this liquid:
Cerebrospinal Fluid
CSF Production
CSF is produced mainly in the choroid plexus, a network of capillaries and ependymal cells located within the ventricular system of the brain....
Fluid Pressure
According to Pascal's law, a fluid at rest will generate equal pressure in all directions. This pressure is measured as a force per unit area, and its magnitude depends on the fluid's specific...
Characteristics of Fluids
Fluids, which include both liquids and gases, are substances that deform continuously under shearing stress. For example, water and oil are liquids with molecules that can...