Diffusion Tensor Magnetic Resonance Imaging in Chronic Spinal Cord Compression
Published on: May 7, 2019
Retrospective distortion correction of diffusion tensor imaging data by semi-elastic image fusion - Evaluation by
Julia Gerhardt1, Nico Sollmann2, Patrick Hiepe3
1Department of Neurosurgery, Klinikum rechts der Isar, Technische Universität München, Ismaninger Str. 22, 81675, Munich, Germany.
Objective:
Diffusion tensor imaging (DTI) based on echo-planar imaging (EPI) can suffer from geometric image distortions in comparison to conventional anatomical magnetic resonance imaging (MRI). Therefore, DTI-derived information, such as fiber tractography (FT) used for treatment planning of brain tumors, might be associated with spatial inaccuracies when linearly projected on anatomical MRI. Hence, a non-linear, semi-elastic image fusion shall be evaluated in this study that aims at correcting for image distortions in DTI.
Patients And Methods:
In a sample of 27 patient datasets, 614 anatomical landmark pairs were retrospectively defined in DTI and T1- or T2-weighted three-dimensional (3D) MRI data. The datasets were processed by a commercial software package (Elements Image Fusion .0; Brainlab AG, Munich, Germany) providing rigid and semi-elastic fusion functionalities, such as DTI distortion correction. To quantify the displacement prior to and after semi-elastic fusion, the Euclidian distances of rigidly and elastically fused landmarks were evaluated by means of descriptive statistics and Bland-Altman plot.
Results:
For rigid and semi-elastic fusion mean target registration errors of 3.03 ± 2.29 mm and 2.04 ± 1.95 mm were found, respectively, with 91% of the evaluated landmarks moving closer to their position determined in T1- or T2-weighted 3D MRI data after distortion correction. Most efficient correction was achieved for non-superficial landmarks showing distortions up to 1 cm.
Conclusion:
This study indicates that semi-elastic image fusion can be used for retrospective distortion correction of DTI data acquired for image guidance, such as DTI FT as used for a broad range of clinical indications.
More Related Videos
Related Concept Videos
Inertia Tensor
The diagonal components of the inertia tensor matrix represent the moments of inertia concerning the principal axes of the object. These primary axes are defined as the axes where the object experiences the least...
Nuclear Fusion
A helium nucleus has a mass that is 0.7% less than that of four hydrogen nuclei; this lost mass is converted into energy during the fusion. This reaction produces about...
Diffusion
Distance Corrections
Elasticity
The elasticity of an object can be described by a stress-strain curve, which represents the relationship between stress...
Power Factor Correction