Related Experiment Video
Updated: Jan 23, 2026

Establishing a Swine Model of Post-myocardial Infarction Heart Failure for Stem Cell Treatment
Published on: May 25, 2020
First-in-man study of Heartech® percutaneous left ventricular partitioning device for treatment of heart failure
Zhengbin Zhu1,2, Jiwei Yu1, Kai Xu3
1Department of Cardiology, Rui Jin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Insights
The Heartech® left ventricular partitioning device (LVPD) shows promise in treating heart failure (HF) post-myocardial infarction. This first-in-man study found the device safe and effective, improving patient outcomes and cardiac function.
Area of Science:
- Cardiovascular Medicine
- Medical Devices
- Regenerative Medicine
Background:
- Chronic heart failure (HF) following myocardial infarction (MI) remains a significant clinical challenge, often leading to adverse left ventricular remodeling.
- Current treatment options for HF post-MI have limitations in addressing the underlying structural changes.
- Percutaneous ventricular restoration (PVR) offers a novel approach to managing HF by reshaping the left ventricle.
Purpose of the Study:
- To evaluate the safety and efficacy of the Heartech® left ventricular partitioning device (LVPD) in a first-in-man (FIM) study.
- To assess the feasibility of the PVR procedure using the Heartech® LVPD in patients with HF post-MI.
- To determine the impact of LVPD implantation on cardiac function, exercise capacity, and quality of life.
Main Methods:
- A prospective, single-arm study involving 16 patients with HF post-MI across three cardiac intervention centers.
- All patients underwent percutaneous implantation of the Heartech® LVPD.
- Safety was assessed by monitoring major adverse cardiovascular and cerebrovascular events (MACCEs), while efficacy was evaluated through echocardiography, cardiopulmonary exercise testing (CPET), 6-minute walk test (6MWT), and the EQ-5D quality of life scale.
Main Results:
- The Heartech® LVPD demonstrated a high device success rate of 93.75% with 100% procedural safety.
- No MACCEs were reported during the short-term follow-up period (36 ± 4.5 days).
- Significant improvements were observed in left ventricular volumes (LVESVi and LVEDVi), left ventricular ejection fraction (LVEF), 6MWT distance, and EQ-5D scores.
Conclusions:
- The Heartech® LVPD is a safe and effective device for percutaneous ventricular restoration in patients with HF post-MI.
- LVPD implantation can lead to significant improvements in cardiac structure and function, as well as enhanced exercise capacity and quality of life.
- This FIM study provides a strong foundation for further investigation into the long-term benefits of the Heartech® LVPD in HF management.
Objective:
This first-in-man (FIM) study aimed to determine the safety and efficacy of the Heartech® left ventricular partitioning device (LVPD) in patients with chronic heart failure (HF) postmyocardial infarction.
Methods:
Sixteen patients were enrolled from three cardiac intervention centers in China. All patients underwent percutaneous ventricular restoration (PVR) procedures with implantation of the Heartech® LVPD. Safety and immediate success rates were recorded. Major adverse cardiovascular and cerebrovascular events (MACCEs) including all-cause mortality, myocardial infarction, stroke, emergent or selective surgery or interventional therapy, renal failure requiring hemodialysis, and major bleeding were recorded. Efficacy features included functional status, echocardiographic characteristics, life quality characteristics including peak oxygen consumption of cardiopulmonary exercise testing (CPET), European five-dimensional health scale (EQ-5D), 6-min walk test (6MWT) at baseline and during follow-up (NCT02938637).
Results:
The device success rate was 93.75% (15 successes in 16 patients) with 100% safety. During follow-up of 36 ± 4.5 days, no MACCEs were found. The left ventricular end-systolic volume index decreased significantly (LVESVi, 72.47 ± 22.77 mL/m2 vs. 50.13 ± 13.36 mL/m2 , p < .001) as did left ventricular end diastolic volume index (LVEDVi, 106.27 ± 28.01 mL/m2 vs. 83.20 ± 16.87 mL/m2 , p = .001). Left ventricular ejection fraction (LVEF, 32.47 ± 6.98% vs. 40.41 ± 6.15, p < .001), 6MWT (383.13 ± 108.70 vs. 453.47 ± 88.24, p < 0.001) and EQ-5D (65.93 ± 11.25 vs. 78.67 ± 8.35, p < .001) improved significantly.
Conclusions:
Heartech® LVPD appeared to be safe and effective for treatment of HF postmyocardial infarction.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Related Concept Videos
Heart Failure II: Pathophysiology
Pathophysiology of Heart Failure
Heart Failure I: Introduction
Heart Failure VI: Adjunct Therapies
Heart Failure Drugs: Diuretics
Heart Failure V: Medical Management