Simulation of H p (10) and effective dose received by the medical staff in interventional radiology procedures
P Ferrari1, F Becker2, Z Jovanovic3
1ENEA-IRP, Radiation Protection Institute, 4 Via Martiri di Monte Sole, 40129 Bologna, Italy.
Insights
Interventional radiology procedures increase radiation exposure for medical staff. This study used Monte Carlo simulations to assess operator radiation doses, finding specific exposure levels normalized to radiation output, crucial for safety protocols.
Area of Science:
- Medical Physics
- Radiological Sciences
- Occupational Health
Background:
- Interventional radiology and cardiology offer alternatives to surgery but increase radiation exposure for patients and staff.
- Operator radiation exposure in cardiology procedures is well-studied, but less is known about interventional radiology staff.
- Accurate dosimetry is essential for managing occupational radiation risks in interventional procedures.
Purpose of the Study:
- To investigate and quantify radiation exposure conditions for medical staff during interventional radiology procedures.
- To evaluate the personal dose equivalent and effective dose for operators relative to radiation output.
Main Methods:
- Utilized Monte Carlo simulations with anthropomorphic phantoms to model radiation scenarios for two operators and a patient.
- Modeled personal dosemeters worn on aprons for comparison with hospital measurements.
- Calculated personal dose equivalent, Hp(10), and effective dose, normalized to the simulated Kerma-Area Product (KAP) value.
Main Results:
- The Hp(10)/KAP value for the primary operator was approximately 10 μSv/Gy·cm² without ceiling shielding, varying with dosemeter position.
- Normalized effective dose ranged from 0.03 to 0.4 μSv/Gy·cm².
- Simulation results showed good agreement with hospital measurements and existing literature data.
Conclusions:
- The study provides valuable dosimetric characterization for interventional radiology procedures.
- Findings support the use of numerical modeling for assessing occupational radiation exposure in these settings.
- Results can inform improved radiation protection strategies for interventional radiology staff.
Abstract:
Interventional radiology and cardiology are widespread employed techniques for diagnosis and treatment of several pathologies because they avoid the majority of the side-effects associated with surgical treatments, but are known to increase the radiation exposure to patient and operators. In recent years many studies treated the exposure of the operators performing cardiological procedures. The aim of this work is to study the exposure condition of the medical staff in some selected interventional radiology procedures. The Monte Carlo simulations have been employed with anthropomorphic mathematical phantoms reproducing the irradiation scenario of the medical staff with two operators and the patient. A personal dosemeter, put on apron, was modelled for comparison with measurements performed in hospitals, done with electronic dosemeters, in a reduced number of interventional radiology practices. Within the limits associated to the use of numerical anthropomorphic models to mimic a complex interventional procedure, the personal dose equivalent, H p (10), was evaluated and normalised to the simulated Kerma-Area Product, KAP, value, indeed the effective dose has been calculated. The H p (10)/KAPvalue of the first operator is about 10 μSv/Gy.cm2, when ceiling shielding is not used. This value is calculated on the trunk and it varies of +/-30% moving the dosemeter to the waist or to the neck. The effective dose, normalised to the KAP value, varies between 0.03 and 0.4 μSv/Gy.cm2. Considering all the unavoidable approximation of this kind of investigations, the comparisons with hospital measurement and literature data showed a good agreement allowing to use of the present results for dosimetric characterisation of interventional radiology procedures.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Blind Procedures
Nursing Interventions I: Taxonomy of Nursing Interventions
A nursing intervention is a treatment or action based on scientific concepts and knowledge from the nursing, behavioral, and physical sciences. Identifying and prioritizing nursing interventions based on the desired outcome...
Radiological Investigation I: X-ray and CT
Receiver Operating Characteristic Plot
Nursing Interventions II: Selecting and Classifying the Nursing Interventions