Evaluation of the accuracy and precision of glucometers currently used in Sri Lanka
J H Liyanage1, H A Dissanayake1, K K K Gamage1
1Diabetes Research Unit, Department of Clinical Medicine, Faculty of Medicine, Colombo, Sri Lanka.
Aims:
Life threatening macrovascular and microvascular complications of diabetes can be minimized by effective glycaemic control. Self monitoring of blood glucose with glucometers is recognized as a cost effective strategy to improve glycaemic control. However accuracy and precision of glucometers will determine the effectiveness of this strategy. We aimed to evaluate accuracy and precision of commonly used glucometers in Sri Lanka.
Materials And Methods:
An observational study was conducted in a tertiary care setting including patients with diabetes and healthy volunteers. Eight commonly used glucometers were used. Blood glucose was measured in 50 participants (16 healthy volunteers, 34 patients with diabetes) in finger prick capillary blood using glucometers and venous blood using standard laboratory methods, and were compared to determine accuracy. Repeated measurements from same glucometer with a single finger prick were made and compared to determine precision.
Results:
Only one glucometer showed insignificant difference to venous plasma glucose values. Only one glucometer met ADA recommended bias of <5%. None of the glucometers fell within the ISO recommendations for accuracy.
Conclusion:
Majority of commonly used glucometers in Sri Lanka do not meet the ADA recommendations and ISO standards for accuracy and precision. However their variations are unlikely to make significant adverse impact on patient management.
More Related Videos
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Accuracy and Precision
Improving Translational Accuracy
Improving Translational Accuracy
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Accuracy and Errors in Hypothesis Testing
In hypothesis testing, the probability of making a Type I error, denoted as α, is commonly set at 0.05. This significance level indicates a 5%...