Related Experiment Video
Updated: Jan 23, 2026

Nondestructive Monitoring of Degradable Scaffold-Based Tissue-Engineered Blood Vessel Development Using Optical Coherence Tomography
Published on: October 3, 2018
Tissue-engineered blood vessel mimics in complex geometries for intravascular device testing
Robert Dalton Chavez1, Sara Leifer Walls1, Kristen O'Halloran Cardinal1
1Department of Biomedical Engineering, California Polytechnic State University, San Luis Obispo, California, United States of America.
Objective:
Intravascular stents are commonly used to treat occluded arteries during coronary heart disease. After coronary stent implantation, endothelial cells grow over the stent, which is referred to as re-endothelialization. Re-endothelialization prevents blood from clotting on the stent surface and is a good predictor of stent success. Blood vessel mimics (BVMs) are in vitro tissue-engineered models of human blood vessels that may be used to preclinically test stents for re-endothelialization. BVMs have been developed in straight geometries. However, the United States Food and Drug Administration recommends that devices intended to treat coronary occlusions be preclinically tested in bent and bifurcated vessels due to the complex geometries of native coronary arteries. The main objectives of this study were to develop and characterize BVMs in complex geometries.
Design:
Bioreactors were designed and constructed so that BVMs could be cultivated in bent (>45°) and bifurcated geometries. Human umbilical vein endothelial cells were sodded onto complex-shaped scaffolds, and the resulting BVMs were characterized for cell deposition. For a final proof of concept, a coronary stent was deployed in a severely angulated BVM.
Results:
The new bioreactors were easy to use and mounting scaffolds in complex geometries in the bioreactors was successful. After sodding scaffolds with cells, there were no statistically significant differences between the cell densities along the length of the BVMs, on the top and bottom halves of the BVMs, or on the inner and outer halves of the BVMs. This suggests cells deposited evenly throughout the scaffolds, resulting in consistent complex-geometry BVMs. Also, a coronary stent was successfully deployed in a severely angulated BVM.
Conclusions:
Bioreactors can be constructed for housing complex-shaped vessels. BVMs can be developed in the complex geometries observed in native coronary arteries with endothelial cells evenly dispersed throughout BVM lumens.
Related Concept Videos
Coordination Number and Geometry
Structure of Blood Vessels
Anatomy of Blood Vessels
Arteries
Arteries circulate oxygenated blood from the heart, except the pulmonary artery, which transports deoxygenated blood to the lungs. Large arteries, such as the aorta,...
Predicting Molecular Geometry
Overview of Blood Vessels
Arteries and Arterioles: Arteries are muscular and elastic vessels that primarily carry oxygenated blood from the heart to body tissues, except for the pulmonary artery, which carries deoxygenated blood. They have thick walls to withstand high pressure and contain a layer of muscle tissue, allowing them to expand or contract as...
Development of Blood Vessels
The initial formation of this system is facilitated by the small amount of yolk present in the ovum and yolk sac. Blood vessels originate from...