Are single plane intraoperative and biplanar postoperative radiographic measurements of acetabular cup position the
Ryan E Harold1, Dimitri Delagrammaticas1, Tyler Keller1
1Department of Orthopaedic Surgery, Northwestern University Feinberg School of Medicine, Chicago, IL, USA.
Purpose:
Supine positioning and the use of fluoroscopy during direct anterior approach total hip arthroplasty (DAA THA) have been reported to improve acetabular component positioning. This study aims to compare 2-dimensional intraoperative radiographic RadLink measurements of acetabular component position with 3-dimensional postoperative 3D SterEOS measurements.
Methods:
Intraoperative fluoroscopy and RadLink (El Segundo, CA, USA) were used to measure acetabular cup position intraoperatively in 48 consecutive patients undergoing DAA THA. Cup position was measured on 6-week postoperative standing EOS images using 3D SterEOS software (EOS Imaging, SA, Paris, France) and compared to RadLink findings using Student's t-test. Safe-zone outliers were identified. We evaluated for measurement difference of > +/- 5°.
Results:
RadLink acetabular cup abduction measurement (mean 43.0°) was not significantly different than 3D SterEOS in the anatomic plane (mean 42.6°, p = 0.50) or in the functional plane (mean 42.7°, p = 0.61). RadLink acetabular cup anteversion measurement (mean 17.9°) was significantly different than 3D SterEOS in both the anatomic plane (mean 20.6°, p = 0.022) and the functional plane (mean 21.2°, p = 0.002). RadLink identified 2 cups outside of the safe-zone. However, SterEOS identified 12 (anatomic plane) and 10 (functional plane) outside of the safe-zone. In the functional plane, 58% of anteversion and 92% of abduction RadLink measurements were within +/- 5° of 3D SterEOS.
Conclusions:
Intraoperative fluoroscopic RadLink acetabular anteversion measurements are significantly different than 3D SterEOS measurements, while abduction measurements are similar. Significantly more acetabular cups were noted to be outside of the safe-zone when evaluated with 3D SterEOS versus RadLink.
Related Concept Videos
Plastic Deformations of Members with a Single Plane of Symmetry
Work Done Over an Inclined Plane
The work done by gravity to move a rigid body, or the work done by an opposing force to move a rigid body against gravity, can be calculated using the center-of-mass framework. It is the line integral of the force of gravity over the path, considered positive if...
Body Planes
The sagittal plane is the plane that divides the body or an organ vertically into right and left sides. If this vertical plane runs directly down the middle of the body resulting in equal division, it is called the midsagittal or median...
Transformation of Plane Stress
Plane Electromagnetic Waves I
The EM field is assumed to be a...
Plane Electromagnetic Waves II