Evaluation of an examination chair to quantify the hip internal rotation angle
Bernd Friesenbichler1, Nicola C Casartelli1,2, Nicola A Maffiuletti1
1Human Performance Lab, Schulthess Clinic, Zurich, Switzerland.
Background:
Deformities of the femoral head-neck junction are associated with limited hip internal rotation, which may lead to femoroacetabular impingement and consequently to hip osteoarthritis. This study compared inter- and intra-observer reproducibility of 3 different methods to quantify hip internal rotation.
Methods:
2 investigators assessed hip internal rotation of 30 asymptomatic participants during 2 separate testing sessions. Internal rotation was assessed by rotating the 90°-flexed hip manually while in a supine position (manual), in an examination chair capable of applying a single load (EC1) and in a newly developed examination chair with 5 load levels (EC2). Inter- and intra-observer reproducibility was compared among methods using reliability (intra-class correlation coefficient, ICC) and measurement error (smallest detectable chance).
Results:
Inter-observer reliability was good for the manual assessment (ICC = 0.83) and excellent for the EC1 and EC2 methods (ICC ⩾ 0.95) with expected measurement errors of 15.9°, 7.1° and 6.8°-14.3°, respectively. Intra-observer reliability was excellent for each method (ICC ⩾ 0.96), although measurement error ranged from 7.6°-11.8° for EC2 and was slightly higher compared to the manual (7.8°) and EC1 (5.9°) methods.
Conclusions:
Reproducibility of EC2 hip internal rotation angle assessment is superior to that of the manual assessment at specific load levels but not to the EC1 method. Future assessment devices need to incorporate a means of precisely producing and quantifying the loads applied to the hip joint in order to improve measurement reproducibility.
More Related Videos
05:02Using Unidirectional Rotations to Improve Vestibular System Asymmetry in Patients with Vestibular Dysfunction
Published on: August 30, 2019
08:40Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs
Published on: April 19, 2024
Related Concept Videos
Chair Conformation of Cyclohexane
The hydrogen atoms linked to carbons are arranged in two different axial and equatorial orientations to achieve this...
Internal Energy
Internal Energy
Internal Receptors
Quantifying Work
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...