A comparative analysis of 3MAX aspiration versus 3 mm Trevo Retriever for distal occlusion thrombectomy in acute
Diogo C Haussen1, Brendan Eby, Alhamza R Al-Bayati2
1Neurology, Neurosurgery and Radiology, Emory University School of Medicine/Marcus Stroke & Neuroscience Center - Grady Memorial Hospital, Atlanta, Georgia, USA.
Background:
Although aspiration and stent retriever thrombectomy perform similarly in proximal occlusions, no comparative series are available in distal occlusions. We aimed to compare the 3 mm Trevo Retriever against the 3MAX thromboaspiration catheter in distal arterial occlusions.
Methods:
A single-center retrospective review of a prospectively maintained databank for patients treated with the 3 mm Trevo stent retriever or 3MAX thromboaspiration as the upfront approach for distal occlusions (middle cerebral artery mid/distal M2/M3, anterior cerebral artery A1/A2/A3 or posterior cerebral artery P1/P2) from January 2014 to July 2018 was performed. The primary outcome was the rate of distal occlusion first-pass reperfusion (modified Thrombolysis in Cerebral Infarction (mTICI) 2b-3).
Results:
Of 1100 patients treated within the study period, 137 patients/144 different arteries were treated with the 3 mm Trevo (n=92) or 3MAX device (n=52). The groups had comparable demographics and baseline characteristics. There was a higher rate of first-pass mTICI 2b-3 reperfusion (62% vs 44%; p=0.03), a trend towards a higher rate of final mTICI 2b-3 reperfusion (84% vs 69%; p=0.05), and lower use of adjuvant therapy (15% vs 31%; p=0.03) with the 3 mm Trevo compared with the 3MAX. The median number of passes (p=0.46), frequency of arterial spasm (p=1.00), rates of parenchymal hematomas (p=0.22)/subarachnoid hemorrhage (p=0.37) in the territory of the approached vessel were similar across the two groups. The 90-day rate of good outcomes (45% vs 46% in the 3 mm Trevo and 3MAX groups, respectively; p=0.84) was comparable. Multivariable regression identified baseline NIH Stroke Scale (NIHSS) score (OR 0.9; 95% CI 0.8 to 0.97; p<0.01) and use of 3 mm Trevo (OR 2.2; 95% CI 1.1 to 4.6; p=0.02) independently associated with first-pass mTICI 2b-3 reperfusion.
Conclusions:
In the setting of distal arterial occlusions, the 3 mm Trevo may lead to higher rates of first-pass reperfusion than direct 3MAX thromboaspiration. Lower NIHSS was found to be associated with improved reperfusion rates as observed in more proximal lesions. Further studies are warranted.
Related Concept Videos
Comparing the Survival Analysis of Two or More Groups
Retrieval
Recall involves accessing information without cues, such as during an essay test, where individuals must retrieve facts and concepts from memory unaided. Another example is remembering the name of a colleague...
ER Retrieval Pathway
The ER uses many checkpoints to prevent the entry of incorrectly folded or a resident protein as cargo onto a transport vesicle. These mechanisms...
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Comparing Experimental Results: Student's t-Test
Comparative Excretory Systems