Comparison of Icare Pro Tonometry and Icare One Tonometry Measurements in Healthy Eyes
Hüseyin Mayalı1, Çağlar Sarıgül1, Emin Kurt1
1Manisa Celal Bayar University Faculty of Medicine, Department of Ophthalmology, Manisa, Turkey
Objectives:
To compare intraocular pressure (IOP) measurements obtained with the Icare Pro tonometer used in clinical practice and the Icare One self-tonometer.
Materials And Methods:
Fifty-two eyes of 52 healthy, right-handed individuals with no prior intraocular surgery or ocular trauma, structural ocular pathology, or systemic disease were evaluated. IOP was first measured using the Icare Pro tonometer. The participants were then told how to use the Icare One tonometer and asked to measure their own IOP. The results were analyzed statistically using SPSS v.24.
Results:
Of the 52 healthy participants, 16 (30.7%) were male and 36 (69.3%) were female. Their mean age was 31.6±6.3 (23-47) years. Mean IOP measured with the Icare Pro was 17.10±6.2 (11.5-25.2) mmHg, and the mean self-measured IOP with Icare One was 14.01±3.4 (7-24) mmHg. When the two methods were compared using Levene’s t-test, there was a significant mean difference of -3.08±0.6 (95% confidence interval: -4.39 -1.78; p<0.001).
Conclusion:
In this study, there was a significant difference between the IOP measurements we made using the Icare Pro and the participants’ self-measured IOP using the Icare One, with the latter being relatively lower. This may be related to the fact that the participants were unfamiliar with using the Icare One. Although the Icare One is a promising tool for glaucoma patients to self-monitor their IOP, further studies are needed.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Muscles of the Eye
Extraocular Muscles
The six extraocular muscles surround the eyeball and control its movements. They are responsible for a wide range of eye motions, including looking up, down, left, right, and...