Clinical Comparison of New Generation 0.071-inch and 0.072-inch Aspiration Catheters
Bradley A Gross1, Ashutosh P Jadhav2, Tudor G Jovin3
1Department of Neurological Surgery, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania, USA; The Stroke Institute, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania, USA.
Introduction:
Three new 0.071-inch and 0.072-inch aspiration catheters have been introduced for stroke thrombectomy. Their comparative efficacy has not been clinically evaluated.
Methods:
We reviewed a prospectively maintained thrombectomy database for cases using 1 of these 3 catheters for proximal large vessel occlusion from September 2018 to February 2019. Clinical and angiographic information was extracted.
Results:
Of 145 thrombectomies performed over the time period, 49 utilized 1 of the 3 new large-bore catheters (React 71, n = 21; Vecta 71, n = 19; Jet 7, n = 9) on the first pass. Mean patient age was 73 years (SD: 15). Mean presenting National Institute of Health Stroke Scale score was 19 (SD: 6, range: 5-33). Clot location was middle cerebral artery first segment in 59% of cases, internal carotid artery in 31%, and basilar in 10%. With or without a stentriever, clot access with the initially selected aspiration catheter was achieved in 47 of 49 (96%) cases. Excluding empiric stentriever usage, the aspiration catheter could be delivered to the clot without needing a stentriever in 87% of cases: 100% with React 71, 93% with Vecta 71, and 43% with Jet 7 (P = 0.002). Final Thrombolysis in Cerebral Infarction score 2b/3 was achieved in 92% of cases overall: 95% with React 71, 89% with Jet 7, and 89% with Vecta 71. Median procedure time was 28 minutes; mean total number of passes was 2.4 (SD: 1.6) with 39% of cases being single-pass cases. These rates did not significantly differ between aspiration catheters.
Conclusions:
In comparing the newest large bore aspiration catheters, similar angiographic efficacy is achieved.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Molecular Comparison of Gases, Liquids, and Solids