Sound analysis of the magnetically levitated left ventricular assist device HeartMate 3™
Per Sundbom1,2, Michael Roth3, Hans Granfeldt2,4
1Department of Medicine and Geriatrics, Höglandet Hospital, Eksjö, Sweden.
Introduction:
The HeartMate 3™ has shown lower rates of adverse events compared to previous devices due to the design and absence of mechanical bearings. For previous devices, sound analysis emerged as a way to assess pump function. The aims of this study were to determine if sound analysis can be applied to the HeartMate 3 in vivo and in vitro and to evaluate an electronic stethoscope.
Method:
Sound recordings were performed with microphones and clinical accessible electronic stethoscope. The recordings were studied in both the time and the frequency domains. Recordings from four patients were performed to determine if in vivo and in vitro recordings are comparable.
Results:
The results show that it is possible to detect sound from HeartMate 3 and the sound spectrum is clear. Pump frequency and frequency of the pulsatile mode are easily determined. Frequency spectra from in vitro and in vivo recordings have the same pattern, and the major proportion (96.7%) of signal power is located at the pump speed frequency ±40 Hz. The recordings from the patients show low inter-individual differences except from location of peaks originating from pump speed and harmonics. Electronic stethoscopes could be used for sound recordings, but the dedicated equipment showed a clearer sound spectrum.
Discussion:
The results show that acoustic analysis can also be performed with the HeartMate 3 and that in vivo and in vitro sound spectrum is similar. The frequency spectra are different from previous devices, and methods for assessing pump function or thrombosis need further evaluation.
Related Concept Videos
Soundness of Cement
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...
Sound Waves
Sound waves are longitudinal in most fluids because fluids cannot sustain any lateral pressure. In solids, however, shear forces help in propagating the disturbance in the lateral direction as well....
Sound Intensity
Speed of Sound in Gases