MRI-guided mid-position liver radiotherapy: Validation of image processing and registration steps
T N van de Lindt1, M F Fast1, S R van Kranen1
1Department of Radiation Oncology, the Netherlands Cancer Institute, Amsterdam, the Netherlands.
Background & Purpose:
To propose a novel mid-position (midP) workflow for MRI-guided liver SBRT and provide a validation of the required midP-MRI generation and registration steps.
Materials & Methods:
The first step of the midP workflow is the generation of a simulation midP-MRI from a 4D-MRI scan using deformable image registration. Next, a planning midP-CT is warped to the midP-MRI to enable planning in the midP-MRI anatomy. For daily MRI-guidance, three different registration methods to the simulation midP-MRI are proposed; (1) 4D rigid registration of all phases of the daily 4D-MRI, (2) 3D rigid registration of the daily midP-MRI, and (3) 3D deformable registration of the daily midP-MRI. The midP-MRI image quality was assessed with respect to 4D-MRI acquisition time, which is related to over-sampling of the data acquisition (i.e. number of dynamics). The deformable registration precision for the midP-MRI generation was validated using the distance discordance metric (DDM). The deformable CT-MRI and daily MRI-MRI registration accuracies were quantified using the 'full circle method'.
Results:
The DDM was 1.5 mm (median) within the liver, independent of the number of dynamics. The root-mean-squared difference between midP-MRIs based on 10 and 60 dynamics was only 5.2%. The full circle CT-MRI deformable registration error had a median 3D vector length of 1.8 mm in the liver. The daily MRI-MRI registration error was submillimeter for all three evaluated methods.
Conclusion:
The feasibility of an MRI-guided mid-position workflow for liver SBRT is supported by the demonstrated high precision of all image processing and registration steps.
Related Concept Videos
Steps in the Modeling Process
Attention is the first necessary component for observational learning. It involves focusing on what the model is doing and saying. For example, if you decide to take a drawing class to enhance your skills, you need to pay close attention to the instructor's words and hand movements. The characteristics of the model significantly...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Reliability and Validity
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...
Position-effect Variegation
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...