Development and evaluation of a patient-specific surgical simulator for endoscopic colloid cyst resection
Vivek P Bodani1,2,3, Gerben E Breimer1, Faizal A Haji4
11Center for Image Guided Innovation and Therapeutic Intervention, The Hospital for Sick Children, Toronto.
Objective:
Endoscopic resection of third-ventricle colloid cysts is technically challenging due to the limited dexterity and visualization provided by neuroendoscopic instruments. Extensive training and experience are required to master the learning curve. To improve the education of neurosurgical trainees in this procedure, a synthetic surgical simulator was developed and its realism, procedural content, and utility as a training instrument were evaluated.
Methods:
The simulator was developed based on the neuroimaging (axial noncontrast CT and T1-weighted gadolinium-enhanced MRI) of an 8-year-old patient with a colloid cyst and hydrocephalus. Image segmentation, computer-aided design, rapid prototyping (3D printing), and silicone molding techniques were used to produce models of the skull, brain, ventricles, and colloid cyst. The cyst was filled with a viscous fluid and secured to the roof of the third ventricle. The choroid plexus and intraventricular veins were also included. Twenty-four neurosurgical trainees performed a simulated colloid cyst resection using a 30° angled endoscope, neuroendoscopic instruments, and image guidance. Using a 19-item feedback survey (5-point Likert scales), participants evaluated the simulator across 5 domains: anatomy, instrument handling, procedural content, perceived realism, and confidence and comfort level.
Results:
Participants found the simulator's anatomy to be highly realistic (mean 4.34 ± 0.63 [SD]) and appreciated the use of actual instruments (mean 4.38 ± 0.58). The procedural content was also rated highly (mean 4.28 ± 0.77); however, the perceived realism was rated slightly lower (mean 4.08 ± 0.63). Participants reported greater confidence in their ability to perform an endoscopic colloid cyst resection after using the simulator (mean 4.45 ± 0.68). Twenty-three participants (95.8%) indicated that they would use the simulator for additional training. Recommendations were made to develop complex case scenarios for experienced trainees (normal-sized ventricles, choroid plexus adherent to cyst wall, bleeding scenarios) and incorporate advanced instrumentation such as side-cutting aspiration devices.
Conclusions:
A patient-specific synthetic surgical simulator for training residents and fellows in endoscopic colloid cyst resection was successfully developed. The simulator's anatomy, instrument handling, and procedural content were found to be realistic. The simulator may serve as a valuable educational tool to learn the critical steps of endoscopic colloid cyst resection, develop a detailed understanding of intraventricular anatomy, and gain proficiency with bimanual neuroendoscopic techniques.
Related Concept Videos
Colloids
Colloids and Suspensions
Colloidal precipitates
Nursing Process for Patient and Caregiver Teaching III: Evaluation and Documentation
Nurses can use several methods to evaluate patient outcomes. For example, oral questions can assess cognitive learning,...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to: