Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
A volume-pressure tail cuff method for hemodynamic parameters: Comparison of restraint and light isoflurane
Nur-Aisha Hem1, James Phie2, Lisa Chilton1
1College of Public Health, Medical and Veterinary Sciences; Centre for Molecular Therapeutics, Australian Institute of Tropical Health and Medicine, James Cook University, 4811 Townsville, QLD, Australia.
Introduction:
A volume-pressure sensor and tail-cuff method for monitoring blood pressure is non-invasive and inexpensive. This method requires animals to be restrained or subjected to anesthesia, but comparative effects of these manipulations on hemodynamic parameters have not been documented.
Methods:
Using a volume-pressure sensor and tail-cuff, we serially measured blood pressure and heart rate in normotensive adult male Lewis rats after light isoflurane-induced anesthesia (5% induction, 1% maintenance) and, following untrained restraint. Blood pressure was recorded until the acquisition of three complete measurements without the range of replicate mean arterial pressures exceeding 15 mmHg (steady-state).
Results:
Averages for the entire series of consecutive measurements indicated that restraint yielded significantly higher systolic and diastolic blood pressure than anesthesia (P < .05), but heart rate was not affected. Following stabilization at steady-state, there were no significant differences in intra- or inter-day hemodynamic values between the restraint and isoflurane groups. The inter-day coefficient of variation for systolic pressure was 13-23% for isoflurane and 9-14% for restraint. Bland-Altman analysis showed wide limits of agreement (±59 mmHg systolic; ±49 mmHg diastolic pressure) between restraint and isoflurane measurements.
Discussion:
Isoflurane caused more variability but there was agreement in BP evaluation by the isoflurane and restraint methods. Using the VPR system, light isoflurane-induced anesthesia and restraint could effectively be used to screen and quantify overt changes in hemodynamic parameters for cardiovascular research utilizing laboratory rodents.