Treatment of Distal Femur Fractures With the DePuy-Synthes Variable Angle Locking Compression Plate
Tyler C McDonald1, Joella J Lambert2, R Miles Hulick1
1Department of Orthopedic Surgery and Rehabilitation, University of Mississippi Medical Center, Jackson, MS.
Objectives:
To determine the failure rate of the DePuy-Synthes variable angle locking compression curved condylar plate (VA-LCP) and quantify failure modes.
Design:
Retrospective review.
Setting:
Level I Trauma Center.
Patients/Participants:
One hundred thirteen patients with 118 OTA/AO classification 33A and 33C distal femoral fractures were included in the study.
Intervention:
Internal fixation using only the DePuy-Synthes VA-LCP plate.
Main Outcome Measurements:
Primary outcomes included mechanical failure rate of the DePuy-Synthes VA-LCP plate in open and closed fractures. Secondary outcomes included overall failure rate of treatment, risk factors for mechanical failure, and the specific location of failure: loss of fixation in the proximal segment, implant failure over the working length, or failure of locking screw fixation distally.
Results:
There were 11 total failures (9.3%) in 118 fractures. Failure rates for the closed and open fracture groups were 5.4% and 15.9%, respectively. Twenty patients (16.9%) required reoperation to promote union. Open fractures (P = 0.00475), the presence of medial metaphyseal comminution (P = 0.037), the length of the zone of comminution (P = 0.037), and plate length (P = 0.0096) were significantly higher in those with implant failure. Most failures (63.6%) were in the working length of the implant.
Conclusions:
The use of the Synthes VA-LCP is a viable option in distal femoral fractures and has an acceptable failure rate and reoperation to promote union rate.
Level Of Evidence:
Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Fluid Pressure over Flat Plate of Variable Width
The pressure distribution on the plate can be calculated by determining the force that acts on a differential area strip of the plate. Thus, the magnitude of the force is equal to the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...