Failure Rates of Base of Thumb Arthritis Surgery: A Systematic Review
Aparna D Ganhewa1, Rui Wu1, Michael P Chae2
1Department of Plastic, Reconstructive and Hand Surgery, Peninsula Health, Frankston.
Purpose:
The purpose of the current review was to estimate failure rates of trapeziometacarpal (TMC) implants and compare against failure rates of nonimplant techniques for surgical treatment of TMC joint (basal thumb joint) arthritis.
Methods:
A systematic review was conducted to identify articles reporting on thumb implant arthroplasty and on nonimplant arthroplasty techniques for treatment of base of thumb arthritis in the English literature. The collected data were combined to calculate failure rates per 100 procedure-years. Failure was defined by the requirement for a secondary salvage procedure. The failure rates between different implant and nonimplant arthroplasty groups were compared directly and implants with higher than anticipated failure rates were identified.
Results:
One hundred twenty-five articles on implant arthroplasty and 33 articles on the outcome of nonimplant surgical arthroplasty of the TMC joint were included. The implant arthroplasty failure rates per 100 procedure-years were total joint replacement (2.4), hemiarthroplasty (2.5), interposition with partial trapezial resection (4.5), interposition with complete trapezial resection (1.7), and interposition with no trapezial resection (4.5). The nonimplant arthroplasty failure rates per 100 procedure-years were: trapeziectomy (0.49), joint fusion (0.52), and trapeziectomy with ligament reconstruction ± tendon interposition (0.23).
Conclusions:
Several implant designs (arthroplasties) had high rates of failure due to aseptic loosening, dislocation, and persisting pain. Furthermore, some implants had higher than anticipated failure rates than other implants within each class. Overall, the failure rates of nonimplant techniques were lower than those of implant arthroplasty.
Type Of Study/Level Of Evidence:
Therapeutic IV.
Related Concept Videos
Review and Preview
Percentiles are a type of fractile that partition data into...
Review and Preview
Random and Systematic Errors
Systematic Sampling Method
Systematic sampling is one of the simplest methods...
Range Rule of Thumb to Interpret Standard Deviation
For instance, the range rule of thumb can be used to find the tallest and the shortest student in a class, given the mean student height and standard deviation. If the mean student height is 1.6 m and the standard deviation, s is 0.05 m, the height...
Propagation of Uncertainty from Systematic Error