Related Experiment Video
Updated: Jan 22, 2026

A Protocol for Constructing a Rat Wound Model of Type 1 Diabetes
Published on: February 17, 2023
Glucose Control in Adults with Type 1 Diabetes Using a Medtronic Prototype Enhanced-Hybrid Closed-Loop System: A
Melissa H Lee1,2, Sara Vogrin1, Barbora Paldus1
1Department of Medicine, University of Melbourne, Melbourne, Australia.
Abstract:
Experience from first-generation closed-loop (CL) systems informs refinements to enhance glucose control and user acceptance. A next-generation prototype enhanced-hybrid CL (E-HCL) system incorporates iterative changes to the Medtronic MiniMed 670G CL system, including automated correction boluses, lower target glucose level, and user enhancements. The aim was to explore safety, system performance, and glucose control using E-HCL in adults with type 1 diabetes. Twelve adults underwent this first in-human feasibility study. After a 1-week run-in using open-loop (OL), E-HCL was activated at the start of a supervised 1-week hotel phase, followed by 3 weeks free living at home. Supervised challenges included two meal interventions (unannounced and late meal bolus) and a sensor calibration intervention. Primary outcome was sensor glucose time-in-range (TIR); OL run-in and E-HCL at home were compared by Wilcoxon signed-rank test. Twelve adults (seven men; median [interquartile range] age 48 [39, 57] years; HbA1c 6.8 [6.2, 7.2]%, 51 [44, 55] mmol/mol; diabetes duration 31 [13, 41] years) completed the protocol. E-HCL resulted in greater TIR (85.3 [79.4, 88.4]% vs. 75.0 [66.6, 83.7]%, P = 0.003) and lower mean sensor glucose (123.0 [119.3, 129.6] mg/dL vs. 143.5 [135.8, 154.5] mg/dL, P = 0.002) than OL. Time spent <70 mg/dL increased using E-HCL (4.4 [3.3, 6.1]% vs. 3.0 [1.8, 3.8]%, P = 0.02) with no difference in time <54 mg/dL (P = 0.64). Time in CL was 99.98 [99.0, 100.0]%. All participants were satisfied using E-HCL. In adults with well-controlled HbA1c levels, a prototype E-HCL resulted in high TIR, few CL exits, and positive user experiences at the expense of increased hypoglycemia (<70 mg/dL). E-HCL represents a positive step in the journey toward optimizing glucose control in people living with type 1 diabetes.
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Diabetes Mellitus: Type 2 and Gestational
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Concepts and Prototypes
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
Hybrid Zones