Repeatability of the Novel Intraocular Pressure Measurement From Corvis ST
Masato Matsuura1,2, Hiroshi Murata1, Yuri Fujino1
1Department of Ophthalmology, The University of Tokyo, Tokyo, Japan.
Purpose:
To assess the repeatability of intraocular pressure (IOP) measured with the Corvis ST (CST) and the Ocular Response Analyzer (ORA).
Methods:
A total of 141 eyes from 141 subjects were studied, including 35 healthy eyes and 106 glaucomatous eyes. All subjects underwent IOP evaluations with Goldmann applanation tonometer, CST, and ORA. With CST, biomechanical corrected IOP (bIOP) was calculated; bIOP is purported to be less dependent on biomechanical properties. For ORA, corneal-compensated intraocular pressure (IOPcc) and Goldmann-correlated IOP (IOPg) were derived. The repeatability of the various IOP values was assessed using the coefficient of variance (CV) and the intraclass correlation coefficient (ICC).
Results:
The CV with bIOP (5.5 ± 3.1: mean ± standard deviation) was significantly smaller than the CVs measured with IOPg (7.3 ± 4.3) and IOPcc (7.2 ± 4.4). ICC values were 0.90, 0.80, and 0.86 with IOPg, IOPcc, and bIOP, respectively.
Conclusions:
The bIOP showed a better prevision and repeatability for IOP measurement.
Translational Relevance:
The bIOP measurement from CST had a better reproducible than IOPcc measurement from ORA.
Related Concept Videos
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Vapor Pressure