Related Experiment Video
Updated: Jan 22, 2026

Biodegradable Magnesium Stent Treatment of Saccular Aneurysms in a Rat Model - Introduction of the Surgical Technique
Published on: October 1, 2017
Multiple pipeline twists encountered during treatment of a symptomatic fusiform ICA aneurysm
Robert W Young1, Matthew T Bender1, Geoffrey P Colby2
1Department of Neurosurgery, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
Abstract:
Pipeline embolisation device (PED) 'twisting' is an intra-operative complication that manifests with the appearance of a 'figure-8' in perpendicular planes on digital subtraction angiography. A twisted PED causes narrowing and/or complete occlusion of the vessel lumen and poses significant risks for thrombus formation and downstream ischaemia. Here, we present a case in which three unique PED implants become twisted during pipeline embolisation of a large fusiform internal carotid artery aneurysm. The twists were remediated by balloon angioplasty and a combination of techniques that allowed the PED to rotate and restore its original axis. Six-month and twelve-month follow-up angiography demonstrated complete aneurysm occlusion with preservation of the parent vessel, proving that proper remediation of PED twisting can still result in successful long-term outcomes.