Related Experiment Video
Updated: Jan 22, 2026

The C-seal: A Biofragmentable Drain Protecting the Stapled Colorectal Anastomosis from Leakage
Published on: November 4, 2010
Protection of colorectal anastomosis with an intraluminal bypass device for patients undergoing an elective anterior
A Reshef1, G Ben-Arie2, I Pinsk3
1Colorectal Unit, Department of Surgery, Soroka University Medical Center, Beersheba, Israel. areshef1@yahoo.com.
Background:
Currently, the only clinically valid method to prevent morbidity and mortality related to colorectal anastomotic leaks is by construction of a protective ileostomy. Intraluminal bypass might also be a possible way to proctect the anastomosis. The aim of the present study was to evaluate the CG-100 intraluminal bypass device for the reduction of anastomosis-related morbidity and stoma creation in cases of rectal resection.
Methods:
A prospective study was conducted on patients having sphincter-preserving rectal resection who were treated with the CG-100 device at Soroka University Medical Center, Beer Sheva, Israel between May 2015 and February 2017. The device was implanted during surgery and removed after 10 ± 1 days. All patients underwent a radiologic leak test with water-soluble contrast prior to removal of the device. Patients were followed for 30 days. Information about adverse events, anastomotic leaks, device usability and tolerance were collected.
Results:
Forty-seven patients participated in the study. Most patients were operated on due to cancer 44 (93.6%). Four (9%) patients received a primary protective stoma on top of the CG-100 device as part of the learning curve of the surgical team and none required a stoma after device removal. Five (9%) serious adverse events were reported, but only 2 (4%) were classified as related to the device. One was a transient enterocutaneous fistula after removal of the device. The second was an asymptomatic radiologic leak in 1 (2.1%) patient which was treated by keeping the device in place and antibiotic treatment for another 10 days without creation of diverting ileostomy.
Conclusions:
CG-100 may provide a safe method for fecal diversion over a newly created anastomosis without the complications related to stoma creation and closure. A larger prospective randomized study in patients originally scheduled to receive diverting stoma is needed to confirm these findings.
Related Concept Videos
Pilot and Numeric Relaying
Zones of Protection
Protective zones are defined by closed dashed lines, containing one or more components. A key characteristic of these zones is the strategic placement of...
Protection of Alcohols
Protection
It defines a protecting group as the masking agent to make the more reactive species inert to a given set of conditions. This concept is depicted via the illustration of liquid flow through different outlets in an assembly of pipes. The analogy helps to understand the role...
Protecting Self-Esteem
Radial System Protection
In a radial system with a fault downstream of the third breaker, ideally, only the third breaker will open, isolating the fault and interrupting the load connected beyond it. The second breaker has a longer delay setting,...
Muscles of the Anterior Neck