Prospective randomized comparison of early versus newer-generation vertebral access devices for kyphoplasty
Falko Schwarz1, Aaron Lawson McLean2, Anna Luisa Steinberg2
1Department of Neurosurgery, Jena University Hospital-Friedrich Schiller University Jena, Am Klinikum 1, 07747, Jena, Germany. falko.schwarz@med.uni-jena.de.
Introduction:
Kyphoplasty is an established method of treating osteoporotic vertebral body compression fractures. In recent years, several techniques to enhance the efficiency and outcomes of this surgery have been developed and implemented in clinical practice. In the present study, we assess the impact of two new access instruments on overall operation time and the administered dose area product in comparison with the standard access instrument used in our clinical practice. The two newer comparator devices have been designed with the intention of streamlining intraoperative workflow by omitting several procedural steps.
Materials And Methods:
This was a single-center prospective randomized trial investigating three distinct access instruments compatible with the Joline Allevo balloon catheter system. Specifically, two newer access devices marketed as being able to enhance surgical workflow (Joline RapidIntro Vertebra Access Device with a trocar tip and Joline SpeedTrack Vertebra Introducer Device with a short, tapered tip) were compared with the older, established Joline Vertebra Access Device from the same firm. Consecutive eligible and consenting patients scheduled to undergo kyphoplasty for osteoporotic vertebral compression fracture refractory to conservative, medical treatment during the period May 2012-August 2015 were randomized to receive surgery using one of the three devices. Besides the use of the trial instruments, all other preoperative, intraoperative and postoperative care was delivered according to standard practice.
Results:
91 kyphoplasties were performed on 65 unique patients during the study period. The median operation time across the three groups was 29 min (IQR 22.5-35.5) with a median irradiation time of 2.3 min (IQR 1.2-3.4). The median patient age was 74 years (IQR 66-80). The groups did not significantly differ in terms of age (p = 0.878), sex (p = 0.37), T score (p = 0.718), BMI (p = 0.285) or the applied volume of cement (p = 0.792). There was no significant difference between the treatment groups with respect to surgical duration (p = 0.157) or dose area product (p = 0.913).
Conclusions:
Although use of the two newer-generation access instruments were designed to involve fewer unique steps per operation, their use was not associated with reduction in surgical duration, irradiation time or dose area product administered compared with the older, established vertebral access device. Care should be taken to evaluate the impact of new instruments on key surgery-related parameters such as surgical duration and radiation exposure and claims made about new instruments should be assessed a structured fashion.
Related Concept Videos
Random Error
Random Variables
Uppercase letters such as X or Y denote a random variable. Lowercase letters like x or y denote the value of a random variable. If X is a random variable, then X is written in words, and x is given as a number.
For example, let X = the...
Randomized Experiments
Simple randomization
Simple...
Muscles of the Vertebral Column
Superficial Layer:
The superficial layer consists primarily of the splenius muscles, which include the splenius capitis and splenius cervicis. These muscles are mainly responsible for the head and cervical spine movements, including extension, rotation, and lateral bending. The splenius capitis...
Articulations of the Vertebral Column
The Sense of Self: Reflected Self-Appraisal and Social Comparison