Related Experiment Video
Updated: Jan 22, 2026

A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
Learning Curve for Robot-Assisted Percutaneous Pedicle Screw Placement in Thoracolumbar Surgery
Jeremy K T Kam1, Calvin Gan2, Stefan Dimou1
1Department of Neurosurgery, Royal Melbourne Hospital, Parkville, VIC, Australia.
Study Design:
Retrospective review of an initial cohort of consecutive patients undergoing robot-assisted pedicle screw placement.
Purpose:
We aimed to evaluate the learning curve, if any, of this new technology over the course of our experience.
Overview Of Literature:
Percutaneous pedicle screws have specific advantages over open freehand screws. However, they require intraoperative imaging for their placement (e.g., fluoroscopy and navigation) and require increased surgeon training and skill with the learning curve estimated at approximately 20-30 cases. To our knowledge, this is the first study that measures the learning curve of robot-guided purely percutaneous pedicle screw placement with comprehensive objective postoperative computed tomography (CT) scoring, time per screw placement, and fluoroscopy time.
Methods:
We included the first 80 consecutive patients undergoing robot-assisted spinal surgery at Melbourne Private Hospital. Data were collected for pedicle screw placement accuracy, placement time, fluoroscopy time, and revision rate. Patient demographic and relevant perioperative and procedural data were also collected. The patients were divided equally into four sub-groups as per their chronological date of surgery to evaluate how the learning curve affected screw placement outcomes.
Results:
Total 80 patients were included; 73 (91%) had complete data and postoperative CT imaging that could help assess that placement of 352 thoracolumbar pedicle screws. The rate of clinically acceptable screw placement was high (96.6%, 95.4%, 95.6%, and 90.7%, in groups 1 to 4, respectively, p=0.314) over time. The median time per screw was 7.0 minutes (6.5, 7.0, 6.0, and 6.0 minutes in groups 1 to 4, respectively, p=0.605). Intraoperative revision occurred in only 1 of the 352 screws (0.3%).
Conclusions:
We found that robot-assisted screw placement had high accuracy, low placement time, low fluoroscopy time, and a low complication rate. However, there were no significant differences in these parameters at the initial experience and the practiced, experience placement (after approximately 1 year), indicating that robot-assisted pedicle screw placement has a very short (almost no) learning curve.
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Heating and Cooling Curves
For instance, the addition of heat raises the temperature of a solid; the amount of heat absorbed depends on the heat capacity of the solid (q = mcsolidΔT). According to thermochemistry, the relation between the amount of heat absorbed or released by a substance, q, and its...
Acid-Base Titration Curves
For a titration carried out for 25.00 mL of...
Curve Equations