Significant differences between specular microscopy and corneal bank endothelial cell counts - a pilot study
Annekatrin Rickmann1, Katrin E Boden1, Silke Wahl1
1Eye Clinic Sulzbach, Knappschaft Hospital Saar, Sulzbach, Germany.
Background:
It was shown recently that endothelial cell count performed by cornea banks overestimates the real number of endothelial cells. The aim of this study was to investigate the internal quality of preclinical ECD in human donor corneas using two widely used methods for endothelial cell counting, transmitted light microscopy used in organ culture tissue bank and clinically used specular microscopy.
Methods:
Twenty human donor corneas that could not be transplanted were included in this analysis. Differences in evaluating endothelial cell density (ECD) and hexagonal endothelial cell ratio (HEX) between clinical specular microscopy (CSM) and corneal bank transmitted light microscope (CBLM) were evaluated as well as differences between automated and manual cell counts.
Results:
Automated CBLM showed a higher ECD of 31.85% compared to automated CSM, while manual CBLM counting is 10.51% higher compared to manual CSM (p < 0.01). Further, higher average ECD values result in a higher difference between CSM and CBLM measurements. The manual CBLM ECDs were significantly higher compared to automated derived ECD from CSM (p < 0.01). However, no systematic bias can be detected when comparing the differences of the measurements with the average ECD measurements of both methods.
Conclusion:
This preclinical pilot study confirmed a significant higher ECD using transmitted light microscopy in organ culture compared to clinical specular microscopy. This indicates that the early rapid decrease of EC universally observed after surgery might be partly artefactual.
Related Concept Videos
Pilot and Numeric Relaying
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Electric Potential and Potential Difference
When a test charge moves from the initial to the final position, the electric potential difference between those positions is defined as the ratio of the change in the potential energy to the charge on the...
Identifying Statistically Significant Differences: The F-Test
Sum and Difference OpAmps
A summing amplifier, or an adder, utilizes an op-amp to merge multiple input signals into a single output signal. When audio signals are introduced into its input channels, the input resistors initiate currents that traverse feedback resistors, resulting in an output voltage. Applying Kirchhoff's current...
Difference Equation Solution using z-Transform
The z-transform facilitates handling delayed signals by shifting the signal in the z-domain, which corresponds to delaying the signal in the time domain, and advancing signals by similarly shifting in the...