Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Clinical performance evaluation of the new hematology analyzer Mindray BC-6000
Yimin Shen1, Jun Cao1, Zhiying Zhou1
1Department of Laboratory Medicine, The First Affiliated Hospital of Soochow University, Suzhou, China.
Introduction:
With the progression of blood analysis technology, hematology analyzers become more complex and diverse. How to choose a superb instrument is a challenge for the laboratories. In the essay, we studied whether the newest BC-6000 hematology analyzer meets the needs of a clinical hematology laboratory.
Methods:
Methods comparison was performed using 350 blood samples from patients between different measurement procedures; the basic analytical performance was also tested, including the throughput, carryover, precision, and linearity in different modes. The flagging performances for blasts, immature granulocytes, and NRBC were compared with manual microscopy.
Results:
There were minimal carryover (<0.30%) and excellent actual blood linearity for all routinely used parameters concerned by the clinicians (R2 ≥ 0.997). Repeatability and reproducibility were satisfactory at all testing levels. The functional sensitivity of leukocytes and platelets in the blood and leukocytes and erythrocytes in body fluid was excellent at the 20% CV level. BC-6000 and XN displayed very high correlations for complete blood count (CBC) parameters and very high consistency for leukocyte differentials and NRBC compared with manual microscopy. BC-6000 showed excellent sensitivity and specificity flagging ability on blasts (82.9% and 82.4%) and NRBC (80.0% and 96.9%). For immature granulocytes, BC-6000 showed excellent sensitivity but common specificity flagging ability (91.7% and 65.6%).
Conclusion:
The clinical performance of BC-6000 is excellent, and the analyzer can provide timely and accurate reporting for most of the small- to large-scale laboratories.
Related Concept Videos
Irritable Bowel Syndrome II: Clinical Features and Diagnostic Evaluation
Irritable Bowel Syndrome (IBS) is classified into subtypes based on the predominant bowel habits as determined by the Bristol Stool Form Scale (BSFS). The subtypes are:
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Nursing Clinical Information System
A Nursing Clinical Information System (NCIS) is a specialized type of healthcare information system tailored to meet the unique needs of nursing practice. It incorporates the principles of nursing informatics to streamline information management and improve the quality of care delivery.
Critical attributes of NCIS include:
Clinical Trials
There are four phases in a clinical trial. A phase one...
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...
Pathophysiology of Cardiac Performance