Related Experiment Video
Updated: Jan 22, 2026

Determination of the Settling Rate of Clay/Cyanobacterial Floccules
Published on: June 11, 2018
Transcapillary Refilling Rate and Its Determinants during Haemodialysis with Standard and High Ultrafiltration Rates
Nicos Mitsides1,2,3, Mauro Pietribiasi4, Jacek Waniewski4
1Division of Cardiovascular Sciences, School of Medical Sciences, Faculty of Biology, Medicine and Health, The University of Manchester, Manchester Academic Health Science Centre, Manchester, United Kingdom, nicos.mitsides3@srft.nhs.uk.
Background:
Achieving euvolaemia using ultrafiltration (UF) during haemodialysis (HD) without inducing haemodynamic instability presents a major clinical challenge. Transcapillary refill is a key factor in sustaining the circulating blood volume (BV) during UF, which is in turn predicted by the rate of refilling. However, absolute plasma refilling rate (PRR), its determinants and variability with UF rate (UFR), have not been reported in the literature.
Method:
We studied paired HD sessions (n = 48) in 24 patients over 2 consecutive mid-week HD treatments. Plasma refilling was measured using real-time, minute-by-minute relative BV changes obtained from the integrated BV monitoring device during UF. A fixed bolus dilution approach at the start of HD was used to calculate absolute BV. The first control HD session was undertaken with a standard UFR required to achieve the prescribed target weight, while during the second study session, a fixed (high) UFR (1 L/h) was applied, either in the first (n = 12 patients) or in the final hour (n = 12 patients) of the HD session. Participants' had their hydration status measured pre- and post-HD using multifrequency bioimpedance (BIS). Blood pressure was measured at 15-min intervals and blood samples were collected at 7 intervals during HD sessions.
Results:
The mean PRR during a standard 4-hr HD session was 4.3 ± 2.0 mL/kg/h and varied between 2 and 6 mL/kg/h. There was a mean time delay of 22 min (range 13.3-35.0 min) for onset of plasma refilling after the application of UF irrespective of standard or high UFRs. The maximum refilling occurred during the second hour of HD (mean max PRR 6.8 mL/kg/h). UFR (beta = 0.60, p < 0.01) and BIS derived pre-HD overhydration index (beta = 0.44, p = 0.01) were consistent, independent predictors of the mean PRR (R2 = 0.49) in all HD sessions. At high UFRs, PRR exceeded 10 mL/kg/h. The total overall plasma refill contribution to UF volume was not significantly different between standard and high UF. During interventions no significant haemodynamic instability was observed in the study.
Conclusion:
We describe absolute transcapillary refilling rate and its profile during HD with UF. The findings provide the basis for the development of UF strategies to match varying PRRs during HD. An approach to fluid removal, which is tailored to patients' refilling rates and capacity, provides an opportunity for more precision in the practice of UF.
Related Concept Videos
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...
Measuring Reaction Rates
Speciation Rates
Related Rates
Concentration and Rate Law
For example, in a generic reaction aA + bB ⟶ products, where a and b are stoichiometric coefficients, the rate law can be written as: