Automated False Alarm Reduction in a Real-Life Intensive Care Setting Using Motion Detection
Carl Muroi1, Sando Meier2, Valeria De Luca2
1Neurocritical Care Unit, Department of Neurosurgery, University Hospital Zurich, 8091, Zurich, Switzerland.
Background:
Contemporary monitoring systems are sensitive to motion artifacts and cause an excess of false alarms. This results in alarm fatigue and hazardous alarm desensitization. To reduce the number of false alarms, we developed and validated a novel algorithm to classify alarms, based on automatic motion detection in videos.
Methods:
We considered alarms generated by the following continuously measured parameters: arterial oxygen saturation, systolic blood pressure, mean blood pressure, heart rate, and mean intracranial pressure. The movements of the patient and in his/her surroundings were monitored by a camera situated at the ceiling. Using the algorithm, alarms were classified into RED (true), ORANGE (possibly false), and GREEN alarms (false, i.e., artifact). Alarms were reclassified by blinded clinicians. The performance was evaluated using confusion matrices.
Results:
A total of 2349 alarms from 45 patients were reclassified. For RED alarms, sensitivity was high (87.0%) and specificity was low (29.6%) for all parameters. As the sensitivities and specificities for RED and GREEN alarms are interrelated, the opposite was observed for GREEN alarms, i.e., low sensitivity (30.2%) and high specificity (87.2%). As RED alarms should not be missed, even at the expense of false positives, the performance was acceptable. The low sensitivity for GREEN alarms is acceptable, as it is not harmful to tag a GREEN alarm as RED/ORANGE. It still contributes to alarm reduction. However, a 12.8% false-positive rate for GREEN alarms is critical.
Conclusions:
The proposed system is a step forward toward alarm reduction; however, implementation of additional layers, such as signal curve analysis, multiple parameter correlation analysis and/or more sophisticated video-based analytics are needed for improvement.
Related Concept Videos
Specialized Care Centers and Settings-II
Rural health centers are specialized care facilities in remote locations with very few medical personnel. The primary care providers who run the centers are mostly Registered Nurse Practitioners. Here, emergency treatment is provided to critically ill or injured patients before they are transferred to the closest hospital. Fortunately, due to advancement in technology, many rural healthcare facilities and professionals have easy access to diagnostic and treatment...
Specialized Care Centers and Settings-I
Daycare centers
They provide several functions. Some facilities care for healthy newborns and children whose parents work, while others are medically focused and care for...
Oxidation-Reduction Reactions
Half-life of a Reaction
False Memories
One primary source of false memories is misattribution, where individuals incorrectly associate external information...
Characteristics of Life

![Technical Aspect of the Automated Synthesis and Real-Time Kinetic Evaluation of [11C]SNAP-7941](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F59557.jpg&w=3840&q=50)