The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial
Masashi Yokose1, Takahiro Mihara1,2, Masahiro Takaya3
1Department of Anesthesiology and Critical Care Medicine, Yokohama City University Graduate School of Medicine, Yokohama, Japan.
Background:
ClearSight is a noninvasive arterial blood pressure monitor, but it remains unknown whether it is affected by the state of perfusion to the fingers. We investigated whether the lower perfusion index (PI) measured with a pulse oximeter, which reflects finger perfusion, would affect the agreement between arterial pressures measured with ClearSight versus those obtained with an arterial catheter.
Methods:
Paired arterial pressure data (ClearSight and arterial catheter-based pressures) and PI values were prospectively obtained from 30 patients undergoing major abdominal surgery. The primary outcome was standard deviation (SD) of the bias (precision) of blood pressure between ClearSight and arterial catheter. The ratio of the adjusted SD of the bias between PI≤1 and PI>1 was calculated using the linear mixed-effects model. The secondary outcomes were the bias and the limits of agreement (LOA) between the two devices (repeated measures Bland-Altman analysis).
Results:
We analyzed 6312 paired data points. The adjusted SD of bias in PI ≤1 compared with those in PI >1 was 1.4-fold (95% confidence interval: 1.3- to 1.4-fold) for systolic arterial pressure, 1.5-fold (95% confidence interval: 1.3- to 1.6-fold) for diastolic arterial pressure, and 1.3-fold (95% confidence interval: 1.2- to 1.5-fold) for mean arterial pressure. The bias (LOA) were as follows: systolic arterial pressure in the PI ≤1 and PI >1 groups, -3.5 (-35.4 to 28.4) mmHg and 2.2 (-19.9 to 24.3) mmHg, respectively; diastolic arterial pressure in the PI ≤1 and PI >1 groups, 13.1 (-5.1 to 31.3) mmHg and 9.0, (-2.6 to 20.6) mmHg, respectively; and mean arterial pressure in the PI ≤1 and PI >1 groups, 8.7 (-11.3 to 28.7) mmHg and 7.6 (-6.2 to 21.3) mmHg, respectively.
Conclusions:
PI ≤1 was associated with a large SD of the bias between the devices. The PI value could be a real-time indicator of ClearSight precision.
More Related Videos
08:51Intra-cardiac Side-Firing Light Catheter for Monitoring Cellular Metabolism using Transmural Absorbance Spectroscopy of Perfused Mammalian Hearts
Published on: May 12, 2019
07:22Quantifying Single Microvessel Permeability in Isolated Blood-perfused Rat Lung Preparation
Published on: June 30, 2014
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.