Vacuum-Sealed Hot Water Bath Immersion for the Preparation of Anatomical and Surgical Cadaveric Bone Models
Published on: December 2, 2022
Fit assessment of two different anatomically contoured distal femur plate sets on cadaveric bones
Ali Utkan1, Volkan İğdir, Aydın Arslan
1Department of Orthopedics and Traumatology, University of Health Sciences, Ankara Numune Training and Research Hospital, 06530 Çankaya, Ankara, Turkey. utkana@yahoo.com.
Objectives:
This study aims to investigate the fitness of two anatomic distal femoral plates with cadaveric femurs and to show whether current plates optimally match each femur.
Materials And Methods:
Two different sets of plates with five, seven and nine shaft holes were applicated on 62 cadaveric femurs. Ball clay was molded onto the entire inner surfaces of the plates and then the plates were fixed to the bones using two self-locking nylon cable zip ties. The volume of ball clay sandwiched in between the plate and bone was calculated and used as a quantitative fit parameter. Data of each plate were analyzed separately.
Results:
Using Double Medical Technology IncorporatedTM plates, the mean plate to bone volumes were calculated as 8.4 mL (range, 5-14 mL), 10.0 mL (range, 6-17 mL), and 13.1 mL (range, 7-25 mL) in five, seven and nine-hole plates, respectively. Using Zimmer Biomet IncorporatedTM plates, the mean volumes were 10.5 mL (range, 6-21 mL), 12.7 mL (range, 7-22 mL) and 16.3 mL (range, 8-30 mL) in five, seven and nine-hole plates, respectively. Within each group, the measurements were significantly correlated positively with femoral length.
Conclusion:
Optimal fit may not be achieved in each femur using current distal femoral plate implant sets. Thus additional sizes of plates should be supplied in the implant sets.
More Related Videos
08:04Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
09:34A Method to Estimate Cadaveric Femur Cortical Strains During Fracture Testing Using Digital Image Correlation
Published on: September 14, 2017
Related Concept Videos
Bones of the Lower Limb: Femur and Patella
Design Example: Analyzing Capacity Contours for Flood Risk Assessment
Topographic Surveying and Contours
Induced-fit Model
Enzymes exhibit substrate specificity, meaning that they can only bind to certain substrates. This is mainly determined by the shape and chemical...
Goodness-of-Fit Test
Inclusive Fitness