A comparative analysis of intensity-based 2D-3D registration for intraoperative use in pedicle screw insertion
Hooman Esfandiari1, Carolyn Anglin2, Pierre Guy3
1Surgical Technologies Lab, Centre for Hip Health and Mobility, Biomedical Engineering, University of British Columbia, H.N. Ho Research Centre, 5th Floor, 2635 Laurel St, Vancouver, BC, V5Z 1M9, Canada. hooman.esfandiari@ubc.ca.
Purpose:
Although multiple algorithms have been reported that focus on improving the accuracy of 2D-3D registration techniques, there has been relatively little attention paid to quantifying their capture range. In this paper, we analyze the capture range for a number of variant formulations of the 2D-3D registration problem in the context of pedicle screw insertion surgery.
Methods:
We tested twelve 2D-3D registration techniques for capture range under different clinically realistic conditions. A registration was considered as successful if its error was less than 2 mm and 2° in 95% of the cases. We assessed the sensitivity of capture range to a variety of clinically realistic parameters including: X-ray contrast, number and configuration of X-rays, presence or absence of implants in the image, inter-subject variability, intervertebral motion and single-level vs multi-level registration.
Results:
Gradient correlation + Powell optimizer had the widest capture range and the least sensitivity to X-ray contrast. The combination of 4 AP + lateral X-rays had the widest capture range (725 mm2). The presence of implant projections significantly reduced the registration capture range (up to 84%). Different spine shapes resulted in minor variations in registration capture range (SD 78 mm2). Intervertebral angulations of less than 1.5° had modest effects on the capture range.
Conclusions:
This paper assessed capture range of a number of variants of intensity-based 2D-3D registration algorithms in clinically realistic situations (for the use in pedicle screw insertion surgery). We conclude that a registration approach based on the gradient correlation similarity and the Powell's optimization algorithm, using a minimum of two C-arm images, is likely sufficiently robust for the proposed application.
Related Concept Videos
Comparing the Survival Analysis of Two or More Groups
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving
To evaluate the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Sound Intensity
Comparative Excretory Systems