A Comparative Study of a Novel Adhesive Bone Conduction Device and Conventional Treatment Options for Conductive
Piotr H Skarzynski1,2,3,4, Anna Ratuszniak1,2, Kamila Osinska1,2
1Institute of Physiology and Pathology of Hearing.
Objective:
To compare the audiological performance with the novel adhesive bone conduction hearing device (ADHEAR) to that with a passive bone conduction (BC) implant and to that with a bone conduction device (BCD) on a softband.
Study Design:
Prospective study in an acute setting, single-subject repeated measure in three situations: unaided, with conventional BCDs (passive implant or on softband), and with the ADHEAR.
Setting:
Tertiary referral center.
Patients:
Ten subjects with conductive hearing loss were evaluated with the ADHEAR. Five of these were users of a passive BC implant (Baha Attract with Baha4); five received a BCD (Baha4) on a softband for test purposes.
Intervention:
Use of non-invasive adhesive bone conduction system for the treatment of conductive hearing loss.
Main Outcome Measures:
Air and bone conduction thresholds, sound field thresholds, word recognition scores in quiet, and speech recognition thresholds in quiet and noise were assessed.
Results:
Users of the passive BC implant received comparable hearing benefit with the ADHEAR. The mean aided thresholds in sound field measurements and speech understanding in quiet and noise were similar, when subjects were evaluated either with the ADHEAR or the passive BC implant. The audiological outcomes for the non-implanted group were also comparable between the ADHEAR and the BCD on softband.
Conclusions:
Based on our initial data, the ADHEAR seems to be a suitable alternative for patients who need a hearing solution for conductive hearing loss but for medical reasons cannot or do not want to undergo surgery for a passive BC implant.
Related Concept Videos
Conduct Disorder
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Resistance and Conductance
Various factors impact the resistance of a conductor. Spiraling in stranded conductors increases their...
Electrical Conductivity
In a practical conductor, an applied electric field may be sustained, causing a flow of electrons, which produce a current. The differential form of the current, the current density, is related to the electric field.
More generally, it is related to the force per unit charge, which involves the...
Theory of Metallic Conduction
In this theory, Newton's second law of motion is used to determine the acceleration of an electron in the presence of an applied electric field. Then, its velocity is expressed via this acceleration.
An electron moves through the crystal, containing positive ions,...