Increasing the discrimination power of rapid evaporative ionisation mass spectrometry (REIMS) in analytical control
Paul Abu-Rabie1, Denver Sheelan1, Alice Laures1
1Medicinal Science and Technology, GlaxoSmithKline R&D, Stevenage, SG1 2NY, UK.
Rationale:
Rapid Evaporative Ionisation Mass Spectrometry (REIMS) has been evaluated as a tool to improve analytical efficiency and add capability in areas within Pharmaceutical Research and Development (Pharma R&D). This article reports the comparison of single MS, and tandem MS/MS REIMS (REIMS and REIMS/MS) methodologies to investigate which mode produces maximum discrimination power for screening applications.
Methods:
Control tissue samples and cell line suspension samples were analysed using optimised REIMS and REIMS/MS to evaluate which technique produced optimal discrimination power for control tissue and cell line identification. The iKnife sampling tool and a prototype 'cell sampler' were utilised for tissue and cell analysis, respectively. The REIMS source was coupled to a hybrid Quadrupole-Time Of Flight (QTOF) mass spectrometer. Multivariate Analysis (MVA) was utilised to evaluate the resulting Mass Spectrometry (MS) data and discriminate between sample types.
Results:
Proof of concept investigations demonstrating that REIMS/MS offered increased MVA discrimination for sample identification, compared with REIMS, is presented for the first time. Control tissue data showed discrimination by timepoint classification over 0-144 h storage after removal from the host. Timepoint discrimination was optimised using REIMS/MS with a collision energy that effectively maximised ion fragmentation. Similar optimisation was observed when REIMS/MS was applied to the identification of cell lines.
Conclusions:
The proof of concept results demonstrate that REIMS/MS can offer advantages over REIMS for control tissue quality screening, and cell line identification applications in Pharma R&D. Further work following this proof of concept investigation is being undertaken to implement the technology for these applications, utilising the optimised REIMS/MS methodology. REIMS/MS will also be used as an optimised tool for other applications.
More Related Videos
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Control of Power Flow
Tandem Mass Spectrometry
Mass Spectrometry: Overview
Mass Spectrometry of Amines