Related Experiment Video
Updated: Jan 22, 2026

Author Spotlight: Innovative Methodology for Implanting and Securing Neural Probes in the Rodent Spinal Cord
Published on: July 12, 2024
Oversensing issues leading to device extraction: When subcutaneous implantable cardioverter-defibrillator reached a
Antoine Noel1, Sylvain Ploux2, Samuel Bulliard1
1Bordeaux University Hospital, Cardio-Thoracic Unit, Pessac, France.
Background:
Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantations are rapidly expanding. However, the subcutaneous detection and interpretation of cardiac signals in S-ICDs is much more challenging than by conventional devices. There is a complete paradigm shift in cardiac signal sensing with subcutaneous signal detection, leading in some cases to oversensing with restricted programming options.
Objectives:
The aim of this single-center study was to quantify and describe cases where recurring oversensing made the extraction of the device necessary.
Methods:
Consecutive patients (n = 108) implanted with an S-ICD in our tertiary referral hospital were considered for analysis. Clinical and remote monitoring data were analyzed.
Results:
The S-ICD had to be explanted in 6 of 108 implanted patients (5.6%) because of refractory oversensing issues: myopotential oversensing, P- or T-wave oversensing, rate-dependent left bundle branch block aberrancy during exercise with R-wave double counting, and R-wave amplitude decrease after ventricular tachycardia ablation leading to noise detection. Seventeen of 108 patients experienced oversensing (15.7%): 9 patients had at least 1 inappropriate charge without a shock (8.3%), 3 patients had at least 1 inappropriate shock (2.8%), and 5 patients had both episodes (4.6%).
Conclusion:
So far, cardiologists have had to deal with transvenous ICD lead fractures, but signal oversensing without correcting programming option could be the equivalent weakness of S-ICDs, despite an adequate screening.
More Related Videos
Related Concept Videos
Ethical Issues
Ethical Concerns in Healthcare:
Issues And Trends In Healthcare Delivery System
Cost Containment
Payment for healthcare services has historically promoted adoption of costly and often unnecessary or inefficient...
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Mnemonic Devices
Acronyms
Acronyms are created by using the initial letters of a series of words to form a new word or phrase. This approach condenses complex information into a single, memorable entity. For example,...
Non-ohmic Devices
Consider a simple circuit consisting of a battery, a diode, and a resistor. A...