Wireless power transfer analysis of circular and spherical coils under misalignment conditions for biomedical
Yellappa Palagani1, Krithikaa Mohanarangam1, Jae Hoon Shim1
1School of Electronics Engineering, College of IT Engineering, Kyungpook National University, Daegu, 702-701, South Korea.
Abstract:
In this paper, a 3-coil inductive WPT system having circular shaped external coil and miniaturized spherical shaped implantable coil is designed and optimized at 13.56 MHz to overcome reduced PTE under misalignment conditions. The external coil is placed at a distance of 10 mm from the implantable coil. The proposed design is referred to as C-S WPT system. To evaluate the performance of the C-S WPT system, a 3-coil inductive WPT system having circular shaped external coil and circular shaped implantable coil is designed and is referred to as C-C WPT system. The impacts of perfect alignment, lateral and/or angular misalignments on the overall PTE of the systems are analyzed and compared. At perfect alignment, the PTE of C-C and C-S WPT systems are 59.8% and 52.6%, respectively. Though the PTE of C-C WPT system is slightly higher than C-S WPT system, there is a sharp decrease in PTE of C-C system for angles greater than 40°. At 90° of angular misalignment, the PTE's of C-C and C-S WPT systems are 0.4% and 5.6% respectively. When the coils are laterally displaced by a distance of 10 mm, the C-C and C-S WPT systems produce PTE of 33.4% and 27.1% respectively. The C-C and C-S WPT systems attain PTE of 1.6% and 5.5% respectively at 10 mm displacement and 50° rotation. The simulated systems are fabricated to analyze the result in real-environment. The measurement results show that the C-S WPT system provides better PTE when compared to C-C WPT system under angular and a combination of angular and lateral misalignments.
Related Concept Videos
Maximum Power Transfer
By substituting the entire circuit with...
The Maximum Power Transfer Theorem
The load connected draws the current, and the circuit delivers the power to the load. The alternating current flowing through the load is determined using the rectangular form of voltages, currents, network impedance, and load impedance. The average power delivered to the load is obtained from the product of the square of current and load resistance.
Spherical Coordinates
Spherical and Cylindrical Capacitor
Conventionally, considering the symmetry, the electric field between the concentric shells of a spherical capacitor is directed radially outward. The magnitude of the field,...
Gravity between Spherical Bodies
This assumption can be proved easily by showing that the expression for gravitational potential energy between a hollow sphere of mass (M) and a point mass (m) is the same as it would be for a pair of extended...
Deformation in a Circular Shaft