Development of Sound Field Audiometry System for Small Audiometric Booths and Comparison of Its Equivalence With
Eun Kyung Jung1, Young Mi Choi1, Eun Jung Kim1
1Department of Otolaryngology-Head and Neck Surgery, Chonnam National University Hospital, Gwangju, Korea.
Objectives:
Sound field (SF) audiometry tests are usually conducted in audiometric booths measuring greater than 2×2 m in size. However, most private ENT clinics carry about 1×1-m-sized audiometric booths, making SF audiometry testing difficult to perform. The aims of this study were to develop an SF audiometry system for use in smaller audiometric booths and compare its performance with traditional system.
Methods:
The newly developed SF audiometry system can yield an SF signal at a distance of about 30 cm from the subject's ears. Its height can be adjusted according to the subject's head height. We compared SF hearing results between the new SF system and the traditional SF audiometry system in 20 adults with normal hearing (40 ears) and 24 adults with impaired hearing levels (38 ears) who wore hearing aids. Comparative parameters included warble tone audiometry threshold, a speech reception threshold (SRT), and a speech discrimination score (SDS). For statistical analysis, paired t-test was used. The equivalence of both SF systems was tested using two one-sided test (TOST) with a margin of 5 dB (normal hearing participants) and 10 dB (hearing aids wearing participants).
Results:
Among participants with normal hearing, warble tone hearing thresholds of 0.5, 1, 2, and 4 kHz, average values of these four frequencies, and SRT were similar between the two systems (all P>0.05). Participants with hearing aids showed similar warble tone threshold and SRT (P>0.05) in both systems except for threshold of 4 kHz (P=0.033). SDS was significantly higher in the newly developed system (P<0.05). TOST results showed equivalent SF audiometry results using either system.
Conclusion:
Audiometric results of the newly developed SF audiometry system were equivalent to those of a traditional system. Therefore, the small SF audiometry system can be used at small audiometric booths present in most private ENT clinics.
Related Concept Videos
Pharmaceutical Equivalents
Equivalent Capacitance
The following strategies are adopted to calculate...
Equivalent Capacitance
Soundness of Cement
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V)...
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...