Performance evaluation of the prototype Abbott Alinity hq hematology analyzer
Yolien Van der Beken1, Annelien Van Dalem1, Guy Van Moer1
1Department of Hematology, Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussel, Brussels, Belgium.
Introduction:
The analytical and clinical performance as well as the workflow efficiency of the novel, prototype Alinity hq hematology analyzer was evaluated in the clinical laboratory of Universitair Ziekenhuis Brussel, Department of Hematology, Brussels, Belgium.
Methods:
Within-run and within-laboratory imprecision, linearity, and carryover were assessed using clinical blood samples and commercial blood products. Four hundred and seventeen samples were selected for method comparison with Abbott CELL-DYN Sapphire, and for flagging performance analysis in comparison with smear review and manual microscopic white blood cell (WBC) differential.
Results:
Within-run and within-laboratory imprecision verification demonstrated low %CV for complete blood count and WBC differential results within the normal ranges (0.1%-10.4%), except for basophil granulocytes. The linearity of the analytical measuring ranges was verified for WBCs, red blood cells, hemoglobin, and platelets. Alinity hq results showed strong agreement with those of CELL-DYN Sapphire. Good correlation was demonstrated with manual WBC differential results, with negative bias for neutrophil (NEU) granulocytes, and positive bias for lymphocytes and monocytes. Blasts were detected with 75% sensitivity and 96% specificity at 1% blast threshold, and 100% sensitivity at 5% blast threshold. Immature granulocyte detection was more sensitive (81% vs 76%, P = 0.086) and specific (88% vs 78%, P = 0.0002) than with CELL-DYN Sapphire. Nucleated red blood cell detection was more sensitive (89% vs 63%, P < 0.001) and just slightly less specific (96% vs 99%, P = 0.0067) than with CELL-DYN Sapphire. Re-run and reflex testing rates were lower with Alinity hq.
Conclusion:
The Alinity hq hematology analyzer is suitable for clinical use.
Related Concept Videos
Concepts and Prototypes
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...
Pathophysiology of Cardiac Performance
Mass Analyzers: Overview
Sources of Self-Esteem II: Performance Feedback
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: