Torsional behaviour of Reciproc and Reciproc blue instruments associated with their martensitic transformation
G C Almeida1, L C Guimarães1, P D Resende2
1Department of Restorative Dentistry, Faculty of Dentistry.
Aim:
To compare the martensitic transformation temperatures and torsional resistance behaviour of classic M-Wire Reciproc and Reciproc blue files (VDW, Munich, Germany).
Methodology:
M-Wire classic Reciproc R25 and Reciproc blue R25 instruments were used. Transformation temperatures were determined by differential scanning calorimetry (DSC-DSC 60, Shimadzu, Kyoto, Japan). Three samples with approximately 20 mg of each system were cooled to -120 °C and then heated to 100 °C and then cooled to -100 °C at a rate of 10 °C min-1 . The maximum torque values and angular deflection until fracture of new instruments (n = 10) were measured according to ISO 3630-1. Results were analysed statistically using the Student's t-test at a significance level of P < 0.05.
Results:
DSC results revealed the absence of martensitic transformation for Reciproc instruments in the temperature range analysed. In both systems, the instruments were composed of a mixture of R-phase and austenite at room temperature. After torsional tests, Reciproc blue had significantly lower values for mean maximum torque at rupture (P < 0.05), although the angular deflection values were significantly higher than the Reciproc classic group (P < 0.05).
Conclusions:
Both Reciproc and Reciproc blue instruments were composed of a mixture of R-phase and austenite. Reciproc Blue instruments had a greater angle of rotation to fracture but a lower torque to failure than M-Wire Reciproc instruments.
Related Concept Videos
Factors Influencing Attraction IV: Reciprocity
Instrument Transformers
Torsion of Noncircular Members
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played...
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...