Background:
The Mindray SAL 8000 is an integrated serum analyzer for photometric, electrochemical, and im-munological assays. The technical, analytical, and workflow performance of the system were evaluated in this study.
Methods:
The technical evaluation was performed using protocols adopted from the guidelines of the China Food and Drug Administration (CFDA). The precision, linearity, interference, and method comparison were carried out according to the Clinical and Laboratory Standards Institute (CLSI) protocols. The verification of carryover and turnaround time were conducted using specimens containing different analytes.
Results:
The technical performance was acceptable for all evaluated aspects. The repeatability and within-labora-tory coefficients of variation (CVs) ranged between 0.22% and 4.23% for routine chemistry and between 1.05% and 6.89% for immunochemistry, respectively. All evaluated analytes exhibited linearity over the ranges claimed by the manufacturer. Significant interferences were observed during low concentration TBIL and P measure-ments due to the presence of lipemia. Method comparisons showed good agreement with the comparison systems and with the correlation coefficients ≥ 0.988 except for anti-HBs (r = 0.812). No significant intra-module and inter-module carryovers were detected. For all the 1,220 samples, 25%, 54%, 63%, 79%, 91%, and 100% samples com-pleted analysis in 16.3 minutes, 30 minutes, 60 minutes, 120 minutes, 180 minutes, and 320 minutes, respectively.
Conclusions:
The Mindray SAL 8000 integrated system achieved optimal technical performance and met most of the criteria regarding analytical performance. The workflow study of the system met the turnaround time (TAT) requirements of laboratories. Therefore, it is a good candidate to be used in medium and large-sized laboratories.
Related Concept Videos
Irritable Bowel Syndrome II: Clinical Features and Diagnostic Evaluation
Irritable Bowel Syndrome (IBS) is classified into subtypes based on the predominant bowel habits as determined by the Bristol Stool Form Scale (BSFS). The subtypes are:
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Chemistry of Carbohydrates
Chemistry of Carbohydrates
What is Organic Chemistry?
Chemistry of the Cell
Water
The polarity of the water molecule and its resulting hydrogen bonding makes water a unique substance with special properties that are intimately tied to the processes of life. Life originally evolved in an aqueous environment, and most of an organism’s cellular chemistry and metabolism occur inside the aqueous contents of the cell’s cytoplasm. Special properties of water are its high heat capacity...