Related Experiment Video
Updated: Jan 22, 2026

Studying Cavitation Enhanced Therapy
Published on: April 9, 2021
Delay multiply and sum beamforming method applied to enhance linear-array passive acoustic mapping of ultrasound
Shukuan Lu1, Renyan Li1, Xianbo Yu1
1The Key Laboratory of Biomedical Information Engineering of Ministry of Education, Department of Biomedical Engineering, School of Life Science and Technology, Xi'an Jiaotong University, Xi'an, 710049, People's Republic of China.
Purpose:
Passive acoustic mapping (PAM) has been proposed as a means of monitoring ultrasound therapy, particularly nonthermal cavitation-mediated applications. In PAM, the most common beamforming algorithm is a delay, sum, and integrate (DSAI) approach. However, using DSAI leads to low-quality images for the case where a narrow-aperture receiving array such as a standard B-mode linear array is used. This study aims to propose an enhanced linear-array PAM algorithm based on delay, multiply, sum, and integrate (DMSAI).
Methods:
In the proposed algorithm, before summation, the delayed signals are combinatorially coupled and multiplied, which means that the beamformed output of the proposed algorithm is the spatial coherence of received acoustic emissions. We tested the performance of the proposed DMSAI using both simulated and experimental data and compared it with DSAI. The reconstructed cavitation images were evaluated quantitatively by using source location errors between the two algorithms, full width at half maximum (FWHM), size of point spread function (A50 area), signal-to-noise ratio (SNR), and computational time.
Results:
The results of simulations and experiments for single cavitation source show that, by introducing DMSAI, the FWHM and the A50 area are reduced and the SNR is improved compared with those obtained by DSAI. The simulation results for two symmetric or nonsymmetric cavitation sources and multiple cavitation sources show that DMSAI can significantly reduce the A50 area and improve the SNR, therefore improving the detectability of multiple cavitation sources.
Conclusions:
The results indicate that the proposed DMSAI algorithm outperforms the conventionally used DSAI algorithm. This work may have the potential of providing an appropriate method for ultrasound therapy monitoring.
Related Concept Videos
Sums of Power
Design Example: Capacitance Multiplier Circuit
The circuit illustrated in Figure 1 below incorporates two op-amps, with the first operating as a voltage follower and the second acting as an inverting amplifier.
Wilcoxon Rank-Sum Test
Sum and Difference OpAmps
A summing amplifier, or an adder, utilizes an op-amp to merge multiple input signals into a single output signal. When audio signals are introduced into its input channels, the input resistors initiate currents that traverse feedback resistors, resulting in an output voltage. Applying Kirchhoff's current...
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Linear Circuits