Posterior wall acetabular fracture fixation: A mechanical analysis of fixation methods
F Pease1, A J Ward1, A J Stevenson1
1Pelvic and Acetabular Reconstruction Unit, Southmead Hospital, North Bristol NHS Trust, Bristol, UK.
Abstract:
Stable, anatomical fixation of acetabular fractures gives the best chance of successful outcome, while penetration of the acetabular articular surface with screws is associated with poor outcomes. Spring plates are an alternative to interfragmentary lag screws when penetration is a concern. A mechanical study comparing fracture stability and construct stiffness of three fixation methods for posterior wall acetabular fractures with transverse comminutions was performed. The three fixation methods tested were a posterior wall rim plate, a posterior wall buttress plate with separate lag screws and a posterior wall plate with two spring plates. Nine samples were tested, three for each fixation method. Two-dimensional motion analysis was used to measure fracture fragment displacement and construct stiffness. After two 6000 cycle-loading protocols, to a maximum 1.5 kN, the mean fracture displacement was 0.154 mm for the rim plate model, 0.326 mm for the buttress plate and 0.254 mm for the spring plate model. Mean maximum displacement was significantly less for the rim plate fixation than the buttress plate (p = 0.015) and spring plate fixation (p = 0.02). The rim plate was the stiffest construct 10,962 N/mm, followed by the spring plate model 5637 N/mm and the buttress plate model 4882 N/mm. Based on data obtained in this study, where possible a rim plate with interfragmentary lag screws should be used for isolated posterior wall fractures as this is the stiffest and most stable construct. When this method is not possible, spring plate fixation is a safe and a superior alternative to a posterior buttress plate method.
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Carbon-dioxide Fixation
Plant Cell Wall
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Bacterial Cell Wall
Composite Masonry Walls