Accuracy of the Ventilator Automated Displayed Respiratory Mechanics in Passive and Active Breathing Conditions: A
Ehab G Daoud1, Reynaldo Katigbak2, Marcus Ottochian3
1Respiratory Care Program, Kapiolani Community College, Honolulu, Hawaii, and John A Burns School of Medicine, University of Hawaii Honolulu, Hawaii. ehab_daoud@hotmail.com.
Background:
New-generation ventilators display dynamic measures of respiratory mechanics, such as compliance, resistance, and auto-PEEP. Knowledge of the respiratory mechanics is paramount to clinicians at the bedside. These calculations are obtained automatically by using the least squares fitting method of the equation of motion. The accuracy of these calculations in static and dynamic conditions have not been fully validated or examined in different clinical conditions or various ventilator modes.
Methods:
A bench study was performed by using a lung simulator to compare the ventilator automated calculations during passive and active conditions. Three clinical scenarios (normal, COPD, and ARDS) were simulated with known compliances and resistance set per respective condition: normal (compliance 50 mL/cm H2O, resistance 10 cm H2O/L/s), COPD (compliance 60 mL/cm H2O, resistance 22 cm H2O/L/s), and ARDS (compliance 30 mL/cm H2O, and resistance 13 cm H2O/L/s). Each scenario was subjected to 4 different muscle pressures (Pmus): 0, -5, -10, and -15 cm H2O. All the experiments were done using adaptive support ventilation. The resulting automated dynamic calculations of compliance and resistance were then compared based on the clinical scenarios.
Results:
There was a small bias (average error) and level of agreement in the passive conditions in all the experiments; however, these errors and levels of agreement got progressively higher proportional to the increased Pmus. There was a strong positive correlation between Pmus and compliance measured as well as a strong negative correlation between Pmus and resistance measured.
Conclusions:
Automated displayed calculations of respiratory mechanics were not dependable or accurate in active breathing conditions. The calculations were clinically more reliable in passive conditions. We propose different methods of calculating Pmus, which, if incorporated into the calculations, would improve the accuracy of respiratory mechanics made via the least squares fitting method in actively breathing conditions.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanism of Breathing II: Expiration
Expiration, or exhaling, is a complex physiological process that begins as the inspiratory muscles begin to relax. This relaxation triggers a series of events that epitomize the efficiency of the respiratory system.
Mechanism of Expiration:
Mechanism of Breathing I: Inspiration
The respiratory system, an essential network for breathing, comprises the conducting and respiratory zones, each playing a crucial role in the overall process of respiration. Let us explore the detailed mechanism of inspiration, or inhalation, which is the first phase of the respiratory cycle.
Pathway of Air during Inspiration
During inspiration, air enters our body through the nose or mouth and moves through the conducting zone,...
Breathing