Back-irradiated and dual-screen sandwich detector configurations for radiography
Anthony R Lubinsky1, Adrian Howansky1, Hao Zheng1
1Stony Brook University, Department of Radiology, Health Sciences Center, Stony Brook, New York, United States.
Abstract:
Recent advances in thin film transistor array technology have enabled the possibility of "back-irradiated" (BI) indirect active-matrix flat-panel imagers (AMFPIs), in which x-rays first expose the scintillator through the optical sensor, and "dual-screen" AMFPIs, in which two scintillating screens are sandwiched around a bidirectional active matrix. We developed a theoretical treatment of such detectors. The theory is used to investigate possible imaging performance improvements over conventional "front-irradiation" (FI) AMFPIs, where the active matrix is opposite the x-ray entrance surface. Simple expressions for the modulation transfer function, normalized noise power spectrum, Swank factor ( ), Lubberts function , and spatial frequency-dependent detective quantum efficiency are derived and used to compute these quantities for a variety of FI, BI, and dual-screen detector configurations. is used as the figure of merit for optimizing and comparing the performance of the various configurations. Large performance improvements over FI single-screen systems are found possible with BI. Further improvements are found possible with dual-screen configurations. The ratio of the thicknesses of the two screens that optimizes DQE is generally asymmetric, with the thinner screen facing the incident flux. The optimum ratio depends on the x-ray attenuation length in the screen.
Related Concept Videos
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configuration of Multielectron Atoms
Stability of Equilibrium Configuration
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...
Configurations of BJT
The common base configuration is noted for its high voltage gain, positioning it as an ideal choice for single-stage amplifier circuits, such as microphone pre-amplifiers. A notable...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...