Repeatability of automated leakage quantification and microaneurysm identification utilising an analysis platform for
Alice Jiang1, Sunil Srivastava1, Natalia Figueiredo1
1The Tony and Leona Campane Center for Excellence in Image-Guided Surgery and Advanced Imaging Research, Cole Eye Instute, Cleveland, Ohio, USA.
Abstract:
Background/aimsUltra-widefield fluorescein angiography (UWFA) provides unique opportunities for panretinal assessment of retinal diseases. The objective quantification of UWFA features is a labour-intensive manual process, limiting its utility. The present study assesses the consistency/repeatability of an automated assessment platform for the characterisation of retinal vascular features, quantification of microaneurysms (MA) and leakage foci in UWFA images.
Methods:
An Institutional Review Board-approved retrospective image analysis study was performed on UWFA images. For each eye, two arteriovenous-phase images and two late-phase images were selected. Automated assessment was performed for retinal vascular features, MA identification and leakage segmentation. Panretinal and zonal assessment of metrics was performed.
Results:
There was a significant correlation between paired time points for retinal vessel area and vessel length on early images (Pearson r=0.92, p<0.0001; Pearson r=0.94, p<0.0001) and late images (Pearson r=0.92, p<0.0001; Pearson r=0.92, p<0.0001, respectively). Panretinal and zonal MA counts demonstrated high repeatability between images (all p<0.0001). Similarly, panretinal leakage area and zonal leakage areas were highly correlated (all p<0.001).
Conclusion:
This automated algorithm demonstrated very strong intrastudy correlation between paired time points in the same phases of the angiogram for quantifying retinal vascular characteristics, MA count and leakage parameters in UWFA images. These findings suggest significant flexibility in the platform for consistency in evaluating metrics over time and is encouraging for longitudinal assessment opportunities.
More Related Videos
11:01Automated Imaging and Analysis for the Quantification of Fluorescently Labeled Macropinosomes
Published on: August 24, 2021
12:02Automated Modular High Throughput Exopolysaccharide Screening Platform Coupled with Highly Sensitive Carbohydrate Fingerprint Analysis
Published on: April 11, 2016
Related Concept Videos
Distribution Reliability and Automation
Methods of Classification and Identification
Agonism and Antagonism: Quantification
To quantify these effects, researchers use a dose-response curve, which provides valuable information about the potency and efficacy of a drug. Potency refers to...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Bioequivalence Experimental Study Designs: Repeated Measures, Cross-Over, Carry-Over, and Latin Square Designs
Qualitative Analysis
For instance, group IV...