Porous tip contact force-sensing catheters for pulmonary vein isolation: performance in a clinical routine setting
Tobias Plenge1, Jan-Hendrik van den Bruck2, Jakob Lüker2
1Department of Electrophysiology, Heart Center, University Hospital Cologne, Kerpener Str. 62, 50937, Cologne, Germany. tobias.plenge@uk-koeln.de.
Purpose:
In catheter ablation of atrial fibrillation (AF), contact force (CF)-sensing catheters with an irrigated tip are used to deliver radiofrequency (RF) energy to the tissue. The ThermoCool® Smarttouch™ Surroundflow catheter (STSF) integrates CF-sensing technology and a new porous tip for advanced external cooling. The aim was to evaluate the performance and safety of STSF in a clinical setting of pulmonary vein isolation (PVI) in comparison with standard contact force-sensing catheter (ST).
Methods:
We assigned consecutive patients (n = 80, prospectively, open-label, non-randomized) with symptomatic AF to either PVI with STSF (n = 60) or ST (n = 20).
Results:
Total ablation time to achieve PVI was significantly shorter in STSF compared to that in ST (STSF, 1556 ± 435 s vs. ST, 1922 ± 961 s; p = 0.045). Ablation time to achieve loss of pace capture of left pulmonary veins was shorter using STSF (left veins, 155 ± 140 s vs. 291 ± 188 s; p = 0.01; right veins, 208 ± 196 s vs. 369 ± 306 s; p = 0.09). Furthermore, administered irrigation fluid was significantly reduced in STSF (241.4 ± 79.6 ml vs. 540.3 ± 229.5 ml; p < 0.01). CF was lower during ablation of left pulmonary veins. One steam pop occurred in STSF, which did not lead to pericardial effusion (vs. no steam pop in ST). The Kaplan-Meier estimate 12-month AF recurrence was 34.3% and 37.7% (p = 0.8).
Conclusions:
Integrating CF technology and the porous tip technology enables effective energy transfer to the tissue resulting in shorter ablation time and less irrigation fluid administration. In our cohort, PVI using the STSF was not associated with an increased complication rate or AF recurrence rate after 12-month follow-up when compared with the ST.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Electromotive Force
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Intermolecular Forces